GMC-I Messtechnik GmbH
31
4.3
Leakage Current Test
This test is required by DIN VDE 0701-0702. It can be used to
test devices, machines and systems for the observance of limit
values for leakage current (contact current).
Leakage current and voltage drop resulting from this current at a
2 k
load are measured and displayed.
Note
If a voltage of greater than 20 V occurs at the test probes,
leakage current measurement is stopped. The 2 k
load
is also deactivated in this case.
Display of measured voltage
U nevertheless continues.
!
Attention!
Interference voltage may not exceed 250 V.
TEST
Select the desired test.
MENU
Open the sub-menu.
4.3.1
Configuring Test Parameters
The
limit value
(maximum leakage current) can be changed.
MENU
Select the desired parameter.
Adjust to the desired value.
4.3.2
Start the Test
START
Note
The leakage current test can only be started with the
START key on the test instrument.