Clamp Multimeter METRACLIP 86
English
Note : - The display is "----" if the input signal is too low (U<5V or I<6A) or if the
frequency is less than 5Hz.
Specific characteristics in MAX/MIN mode
in THD (from 10Hz to 1kHz):
•
Uncertainties: add 1% R to the values in the tables above.
•
Capture time of the extrema: approximately 100ms
4.2.22 Indication of order of the phases
Frequency range
47Hz to 400Hz
Acceptable voltage range
50V to 1,000V
Duration of acquisition of the reference period
≤
500ms
Duration of validity of the reference period
information
approximately 10s at 50Hz
approximately 2s to 400Hz
Duration of acquisition of the measurement period
+ display of the order of the phases
≤
500ms
Acceptable phase unbalance
±
10
Acceptable amplitude unbalance
20%
Acceptable level of harmonics in voltage
10%
4.3
ENVIRONMENTAL CONDITIONS
Environmental conditions
in use
in storage
Temperature
-20 C to + 55 C
-40 °C to + 70°C
Relative humidity (RH):
≤90% at 55°C
≤90% up to 70° C