Clamp Multimeter METRACLIP 86
Note (1) -The display indicates “+OL" above 1800A and "-OL" below –1800A in
REL mode. The “-“ and “+” signs are managed (polarity).
Note (2) - The residual current at zero depends on the remanence; It can be
corrected by the “DC zero” function of the HOLD key.
4.2.5
AC current measurement
Measurement range
(2)
0.15 A to
59.99 A
60.0 A to
599.9 A
600 A (1)
Specified
measurement range
0 to 100% of the measurement range
Uncertainties
± (1% R + 10 pt)
± (1% R +3 pt)
Resolution
0.01A
0.1A
1A
Note (1) - The display indicates "OL" above 900A in PEAK mode. The "-" and "+"
signs are not managed.
- Bandwidth in AC = 3 kHz
Note (2) - Any value between zero and the min. threshold of the measurement
range (0.15V) is forced to “----“ on the display.
- Residual current at zero <150mA.
4.2.6
AC+DC intensity measurement
Measurement range
(2)
0.15 A to
59.99 A
60.0 A to
599.9 A
AC: 600 A
DC or PEAK: 600 A to
900 A (1)
Specified
measurement range
0 to 100% of the measurement range
Uncertainties (2)
(zero corrected)
± (1% R+10
pt)
± (1% R +3 pt)
Resolution
0.01A
0.1A
1A
Note (1) - In DC, the display indicates "+OL" above +1800A and "+-OL" above -
1800A in REL mode. The "-" and "+" signs are managed (polarity).
- In AC and AC+DC, the display indicates "+OL" above 900A in
PEAK mode. The "-" and "+" signs are not managed.
- Bandwidth in AC = 3 kHz
Note (2) - In AC, any value between zero and the min. threshold of the
measurement range (0.15A) is forced to "----" on the display.
- Residual current at zero:
37