LX-80-15 10 Hz User Manual
38
LX-80-15 10 Hz Oceanographic Radar Sensor
Table 9. Writing Data to the Sensor
Fun
Register
Address
Data
Length
Data Range
Details
0X06
0x0003
2 bytes
0
→
9600
1
→
38400
2
→
57600
3
→
115200
4
→
1200
5
→
19200
RS-232 baud rate
0x0004
2 bytes
1 – 255
Modbus ID
0x0005
2 bytes
0
→
9600
1
→
38400
2
→
57600
3
→
115200
4
→
1200
5
→
19200
RS-485 baud rate (Modbus)
0x0006
2 bytes
0
→
no parity, 1 stop bit
1
→
no parity, 2 stop bits
2
→
odd parity, 1 stop bit
3
→
odd parity, 2 stop bits
4
→
even parity,1 stop bit
5
→
even parity, 2 stop bits
RS-485 parity and stop bits
0x0007
2 bytes
1 – 65535
SDI-12 ID
0x000B
2 bytes
1 – 1000
default: 100
Filter length
0x000C
2 bytes
1 – 65535
default: 0
Minimum spectrum amplitude
threshold, used for detecting
peaks
0x000D
2 bytes
0 – device range
default
→
0
Active zone minimum
0x000E
2 bytes
0 – device rang
default
→
device range
Active zone maximum
0x000F
2 bytes
0 – device range
default
→
device range
4-20 mA minimum value in
selected unit
0x0010
2 bytes
0 – device range
default
→
device range
4-20 mA maximum value in
selected unit
0x0011
2 bytes
0 – device range
Level measurement offset
0x0012
2 bytes
0 – 1
IR filter constant
Содержание LX-80-15