3. Check that the distance between the instrument and the water surface is
within the instrument’s operating range.
Use a tape measure or a similar tool to measure the distance between the
instrument’s antenna and the water surface. If the distance to the water surface is
greater than the instrument’s operational range, the instrument will not be able to
measure the distance and the water level.
4. Check that the distance between the instrument and the water surface is
within configured active zone.
Connect the instrument to the Geolux Instrument Configurator PC application
over RS-232 connection. Check the active zone parameters – if the distance to the
actual water surface is outside of the predefined active zone, the instrument will
not be able to report the correct distance. Modify the active zone parameters so
that the actual distance to the water surface is within the active zone.
5. Update device firmware
Connect the instrument to the Geolux Instrument Configurator PC application over RS-
232 connection. Check the active zone parameters – if the distance to the actual water
surface is outside of the predefined active zone, the instrument will not be able to report
the correct distance. Modify the active zone parameters so that the actual distance to
the water surface is within the active zone.
6. Re-calibrate the device
If none of the above mentioned suggestions have helped resolve the problem,
perform device re-calibration. The instrument has the capability to perform self-
calibration. Normally there is no need to re-calibrate the instrument after its initial
calibration in the factory, but in rare cases, re-calibration can resolve the operating
problems of the device. To re-calibrate the instrument, connect it over RS-232 to
the Geolux Instrument Configurator PC application, and then, under
Settings
, click
on the
Recalibrate radar
button.
Firmware upgrade
has failed, or
firmware upgrade
gets stuck on
“Erasing flash”, or
there is no reading
after performing
the firmware
upgrade.
1. Repeat the firmware upgrade procedure
If the firmware upgrade fails, gets stuck on
Erasing flash
, or if after the firmware
upgrade it is not possible to get any readings from the instrument, power off
the device and close the Geolux Instrument Configurator application. After that,
power the instrument back on, re-establish the RS-232 connection to the Geolux
instrument Configurator PC application, and perform the firmware upgrade
procedure again.