LX-80-15 10 Hz User Manual
54
LX-80-15 10 Hz Oceanographic Radar Sensor
4. Check that you are using the correct SDI-12 device ID
The default SDI-12 device ID is A. The device ID can be changed by connecting the
instrument to the PC application (over RS-232 connection).
Connect the instrument to the PC using RS-232 connection. Open the Geolux
Instrument Configurator application and establish a connection between the
instrument and the PC. Then check the
SDI-12 ID
parameter and make sure that it
is the same as the device ID used in issued SDI-12 requests.
5. Make sure that there are no devices with the same device ID connected on
the bus.
SDI-12 allows to have multiple devices connected on the same bus simultaneously.
Each device must have a unique device ID assigned, so that the SDI-12 master
device (datalogger) can distinguish between the devices. If two or more devices
are assigned the same device ID, a bus conflict will happen and prohibit the master
to correctly communicate with the slave devices. To resolve this problem, change
the instrument’s SDI-12 device ID to a unique value through the Geolux Instrument
Configurator PC application.
The instrument
consumes too
much power while
operating in SDI-
12 mode (it does
not reduce power
between two
SDI-12 Measure!
Commands)
1. Switch the instrument from operating mode to standby mode.
Establish a connection to the device over RS-232 connection from the Geolux
Instrument Configurator PC application. Change the Operation mode parameter
from Operating mode to Standby mode.
The 4-20 mA
output is not correct
1. Make sure that the cable is properly connected to the instrument
Check that the cable M12 circular connector is firmly connected and screwed
to the instrument. It is not sufficient to simply attach the cable connector to the
instrument connector, the cable connector must be screwed into the instrument
connector.
2. Check power supply
Make sure that the power is being supplied to the instrument. A direct-current
voltage, between 9 and 27 Volts must be connected to brown (+) and white (-)
wires of the instrument cable. The power supply must be able to deliver at least
500 mA of current.