LX-80-15 10 Hz User Manual
28
LX-80-15 10 Hz Oceanographic Radar Sensor
Modbus stop bits
- Stop bits are added to the end of each data byte transferred over serial
communication, to allow pause between two bytes. One or two bits may be used. The default
setting on most Modbus RTU devices is one stop bit, but some dataloggers may require that the
instrument is configured to use two stop bits.
SDI-12 ID
- The SDI-12 device ID to be used on SDI-12 interface. In SDI-12 request/response
protocol, this ID will be used to define the instrument address, and the instrument will respond only
to requests with matching ID.
4 – 20 mA min.
- To configure the 4 – 20 mA output range, the minimum measured value which
will correspond to 4 mA analog output needs to be set. The value is set in the currently configured
measurement unit. Example: if values measured by the instrument are expected to be within the range
of 700 mm to 5000 mm, it is recommended to configure the minimum value to slightly below 700 mm
(for example 500 mm). Alternatively, if the resolution is not critical, then minimum value for 4 – 20 mA
output can be left to the instrument minimum of 0 mm.
4 – 20 mA max.
– To configure the 4 – 20 mA output range, the maximum measured value
which will correspond to 20 mA analog output needs to be set. The value is set in the currently
configured measurement unit. Example: if values measured by the instrument are expected to be
within the range of 700 mm to 5000 mm, it is recommended to configure the maximum value to
slightly above 5000 mm (for example 6000 mm). Alternatively, if the resolution is not critical, then
maximum value for 4 – 20 mA output can be left to the instrument maximum.
Change processing parameters:
Filter type
- Changing the type of filter used for smoothing measured data.
No filter
- No filtering is used and the raw measurements are reported.
IIR
- Infinite-Impulse Response filter is used to smooth the data. When compared to
moving average filter, IIR filter reacts more quickly to initial change in the data, but it
takes longer for the smoothed value to reach the new measurement. The use of IIR
filter is discouraged for general applications. The IIR constant can be configured
separately.
Moving average
- The moving average filter calculates the average value of a number of
raw measurements. The length for the moving average filter is configured separately
through Filter length parameter.
Median
- The median filter finds the median value from a number of raw measurements.
The length for the median filter is configured separately through
Filter length
parameter.
#set_analog_min=<value>
#set_analog_max=<value>
#set_modbus_stopbits=1
#set_modbus_stopbits=2
#set_sdi_id=<0-61>