Moxi V – Tray insert replacement procedure
Background
The Moxi V has a tray design that allows users to address the occasional need to clean the
system cassette tray of adhesive build-up. The adhesive build-up can eventually lead to
increased resistance to the insertion and removal of cassettes from the system. When
resistance to inserting/removing cassettes becomes prohibitive, users can simply remove a
tray insert (that has the adhesive build-up), replacing it with a new insert (or an old one that
has had the adhesive cleaned off).
Ordering Information
•
Complimentary replacement inserts (up to 5 per instrument, per year) will be provided for
systems under warranty.
•
Additional inserts, or inserts for non-warranty systems, may be purchased using product
code MXA090.
Removing an existing insert
IMPORTANT: Make sure to unplug Moxi V USB/Power cable from system before
starting this procedure. Leaving it in place can result in damage to both the cable
and the system when the system is being tilted back (connectors can break).
1. Open the cassette door all the way.