Moxi V
™
User Guide
Page 28
The data scatter plots/histograms and associated system-set gates can be viewed and adjusted
by the user. It is recommended, particularly for the first test in a sequence of experiments, that
these gates be checked by the user. The scatter plot is accessed by touching the “Next” icon
(image below/right). The Pin Diode (Propidium Iodide, viability) vs size scatter plot is display
(above/middle). The placement of the size gates (blue markers, vertical lines, two top dashed
red arrows in image below/right) and fluorescence gate (horizontal red line, two lower dashed
red arrows in image below/right). The gates, size scale, and axis display can all be adjusted as
described in the “Manual Gating – Fluorescence vs Size Gating” section. Briefly, gates are
moved by touching and dragging the markers (blue for size gates, red for fluorescence gate,
yellow for noise gate). The gate mode (size, fluorescence, noise) can be toggled by touching
the red icon at the top/right of the plot.