Moxi V
™
User Guide
Page 21
on the system “Auto Gate Setting” setting (see “settings” section).
Note: All test results are
ultimately based on the gating of the raw data displays. Even with the simplified tests, the user
always has the option to view/review the raw data and adjust the gates if needed.
The raw data can be displayed in two main formats: 1.). Histograms (one channel display –
either size or PIN Diode fluorescence) – Values for each event are binned across a 1000 point
histogram and displayed. 2) Fluorescence vs. Size scatter plots (all fluorescent assays) – For
each event, the recorded PIN diode millivolt level (range 0-10,000mv, corresponding to
fluorescence intensity on that channel) is plotted against the recorded cell size (effective
diameter in microns).
Detailed labeling for each of the two types of display are provided on the two pages/images
below. After that, an explanation of how to read the key statistics (e.g. counts, sizes, and
percentages) from those data displays is provided.