
80224_MHW_650 L - 1250 L_07-2018_ENG - 68
4. CONFIGURATION
4.14.1. U.CAL - Selecting the user calibration
Acronym
Scrolling message
Submenu
Attributes
U.CAL
USER CALIBRATION TYPE
US.CAL
R W
The parameter shows and sets the parameter, input or output to which calibration will be applied.
Unit of measurement
: -
Options
:
NONE
= No calibration
P.DAYS
= Reset partial day count
I.MAIN
= Calibration of custom main input (selected with parameter tyPE on I.MAIN menu)
OUT.A1
= Custom re-transmission output calibration (selected with parameter t.o.A1 in menu
OUT.AN)
4.14.2. FI.CAL - Resetting the factory calibration
Acronym
Scrolling message
Submenu
Attributes
FI.CAL
FACTORY CALIBRATION
US.CAL
R W
The parameter shows and sets resetting of the factory calibration.
This operation can be done only for inputs and outputs, if U.CAL corresponds to I.MAIN, OUT.A1.
Unit of measurement
: -
Options
:
NO
= Keep user calibration
YES
= Reset factory calibration
4.14.3. C.LOW - Calibrating minimum current / voltage
Acronym
Scrolling message
Submenu
Attributes
C.LOW
US.CAL
R W
The parameter appears if you are calibrating a custom I.MAIN input in current or voltage.
To calibrate:
•
apply the current or voltage value corresponding to minimum scale value to the selected input;
•
press the
key to acquire the calibration value.
Unit of measurement
: -
Options
: -
4.14.4. C.HIGH - Calibrating maximum current / voltage
Acronym
Scrolling message
Submenu
Attributes
C.HIGH
US.CAL
R W
The parameter appears if you are calibrating a custom I.MAIN input in current or voltage.
To calibrate:
•
apply the current or voltage value corresponding to maximum scale value to the selected input;
•
press the
key to acquire the calibration value.
Unit of measurement
: -
Options
: -