Electromagnetic immunity
Table 2-2: Electromagnetic immunity (Part 1)
Guidance and manufacturer’s declaration – electromagnetic immunity.
The Vscan Air CL probe is intended for use in the electromagnetic environment below.
The customer or the user of the Vscan Air CL should assure that it is used in such an
environment.
Immunity test
EN/IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic
discharge (ESD)
EN/IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30%.
Electrical fast
transient/ burst
EN/IEC 61000-4-4
±2 kV for
power-supply lines
±1 kV for input/
output lines
±2 kV for
power-supply
lines
±1 kV for input/
output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
EN/IEC 61000-4-5
±1 kV line(s) to
line(s)
±1 kV line(s) to
line(s)
Mains power quality
should be that of a
typical commercial or
hospital environment.
Содержание Vscan Air
Страница 5: ...Vscan Air User Manual i 3 Direction GP092020 1EN Rev 18 This page intentionally left blank...
Страница 14: ...i 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 212: ...Vscan Air Configuration for iOS and Android 4 60 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 324: ...Using Vscan Air CL 5 112 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 370: ...Vscan Air Maintenance 6 46 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 382: ...Privacy and Security 7 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 406: ...Index 2 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 407: ......