Vscan Air Maintenance
6-4
Vscan Air
–
User Manual
Direction GP092020-1EN
Rev 18
Quality assurance
A good Quality Assurance Evaluation program consists of
periodic systematic actions that provide the user with adequate
confidence that their diagnostic ultrasound system will produce
consistently high-quality images and quantitative information.
Therefore, it is in the best interests of every ultrasound user to
routinely monitor equipment performance.
Typical Tests to Perform
Quality assurance measurements provide results relating to
system performance. Typically. these are:
• Axial
Measurement
Accuracy
•
Lateral Measurement Accuracy
•
Axial and Lateral Resolution
• Penetration
•
Functional and Contrast Resolution
•
Gray Scale Photography.
With these tests, a performance baseline can be set at
installation with the phantom in your department. Future test
results can be compared to the baseline in order to maintain a
record of system performance trends.
Содержание Vscan Air
Страница 5: ...Vscan Air User Manual i 3 Direction GP092020 1EN Rev 18 This page intentionally left blank...
Страница 14: ...i 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 212: ...Vscan Air Configuration for iOS and Android 4 60 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 324: ...Using Vscan Air CL 5 112 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 370: ...Vscan Air Maintenance 6 46 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 382: ...Privacy and Security 7 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 406: ...Index 2 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 407: ......