Cleaning and disinfection
Vscan Air
–
User Manual
6-11
Direction GP092020-1EN
Rev 18
Cleaning using a cleaning solution
1. Prepare a basin with enzymatic cleaner per the
manufacturer’s instructions.
2. Immerse probe in the cleaning solution and ensure no air
bubbles are trapped.
3. Use a soft nylon bristle brush to clean the probe. Do Not use
the brush on the probe lenses as this can damage the soft
acoustic lens.
4. Ensure the probe remains in the cleaning solution for the
minimum contact time listed on the enzymatic cleaner label.
Make sure to follow manufacturers’ recommendations.
5. Visually inspect the probe and probe components for soil.
Repeat steps 4 - 6 until all visible soil has been removed
from the surface of the probe.
6. Thoroughly rinse the probe to remove traces of the cleaner
solution. Follow the cleaner manufacturers’ instructions.
7. Visually inspect the device in a well-lit area to ensure all
surfaces are free from residual cleaning solution. Repeat
Step 6 if visible cleaning solution is observed.
8. Thoroughly pat dry the probe using a clean low-lint soft, dry
disposable cloth or wipe. Do not use abrasive paper
products.
Содержание Vscan Air
Страница 5: ...Vscan Air User Manual i 3 Direction GP092020 1EN Rev 18 This page intentionally left blank...
Страница 14: ...i 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 212: ...Vscan Air Configuration for iOS and Android 4 60 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 324: ...Using Vscan Air CL 5 112 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 370: ...Vscan Air Maintenance 6 46 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 382: ...Privacy and Security 7 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 406: ...Index 2 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 407: ......