Package contents
Vscan Air
–
User Manual
3-7
Direction GP092020-1EN
Rev 18
How to charge the Vscan Air CL (continued)
NOTE:
The Vscan Air CL will charge while being powered off. If the
probe is placed onto the charging pad while being turned on
it will Initiate a shutdown within 30 seconds after detecting
that charging has started.
NOTE:
Both LEDs will be blinking in case of any kind of charging
issue. This is most likely caused by the probe and battery
being too warm start charging. Please let the probe cool off
and check that probe starts charging indicated by steady
LED light in both LEDs.
NOTE:
The probe charge status will be monitored at regular
intervals if probe is left fully charged on the charging pad.
Charging will be restarted as required to keep the probe fully
charged while placed on the wireless charging pad.