Safety
2-24
Vscan Air
–
User Manual
Direction GP092020-1EN
Rev 18
Maximum probe temperature
Maximum probe temperature
The table below indicates the maximum probe temperature.
NOTE:
Measurement uncertainty and probe variation: 2.0 °C.
NOTE:
The probe will stop scanning, if exceeding an applied part
surface temperature of 43°C
NOTE:
Lens temperature is measured under conditions per IEC
60601-2-37, Ed2.1
•
Thermocouple was placed at the geometric center of the
lens.
•
Thermal phantom at not less than 33°C or in the range
20-33°C for external probes. Maximum Vscan Air CL
temperature rise is measured and added to 33°C. Maximum
Vscan Air CL temperature (Simulated use) is <43°C.
•
With Vscan Air CL transmitting in air, temperature rise is
measured and added to 23°C. Maximum Vscan Air CL
temperature (Still air) is <50°C. Lens temperature is
monitored for 30 minutes.
NOTE:
Vscan Air CL placed in contact with a thermal phantom
made of tissue-mimicking material is referenced in
IEC60601-2-37, Ed2.1
•
Auto-freeze capability is disabled during these
measurements.
Table 2-7: Maximum Vscan Air CL temperature
Probe
Max Temp (°C)
(Simulated use)
Max Temp (°C)
(Still air)
Vscan Air CL – Curved array transducer
(for deep scanning)
40.5
45.3
Vscan Air CL – Linear array transducer (for
shallow scanning)
41.1
44.8
Содержание Vscan Air
Страница 5: ...Vscan Air User Manual i 3 Direction GP092020 1EN Rev 18 This page intentionally left blank...
Страница 14: ...i 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 212: ...Vscan Air Configuration for iOS and Android 4 60 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 324: ...Using Vscan Air CL 5 112 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 370: ...Vscan Air Maintenance 6 46 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 382: ...Privacy and Security 7 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 406: ...Index 2 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 407: ......