Vscan Air Maintenance
6-2
Vscan Air
–
User Manual
Direction GP092020-1EN
Rev 18
System care and maintenance
Overview
The Vscan Air CL requires regular care and maintenance to
function safely and properly.
The expected service life of the Vscan Air CL is 5 years.
To ensure that the Vscan Air constantly operates at maximum
efficiency, we recommend that the following procedures be
observed as part of the customer’s internal routine maintenance
program.
NOTE:
The user must ensure that safety inspections are performed at
least every 12 months according to the requirements of the
patient safety standard EN/ES/IEC 60601-1 and its national
deviations or according to the requirements of the IEC62353
standard, medical electrical equipment – recurrent test and test
after repair of medical electrical equipment.
Special attention shall be paid to the following:
•
Safety related marking / labels are legible and complete for
the Vscan Air probe
•
The integrity of the probe e.g. is there any observed
damage to the Vscan Air probe
•
Assess the accessories for damage (AC adapter, USB
cable, wireless charger pad), and
•
Assess if the required documentation is available and
reflects the current revision and/or configuration of the
Vscan Air
CAUTION
Only trained persons should perform the safety inspections.
Содержание Vscan Air
Страница 5: ...Vscan Air User Manual i 3 Direction GP092020 1EN Rev 18 This page intentionally left blank...
Страница 14: ...i 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 212: ...Vscan Air Configuration for iOS and Android 4 60 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 324: ...Using Vscan Air CL 5 112 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 370: ...Vscan Air Maintenance 6 46 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 382: ...Privacy and Security 7 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 406: ...Index 2 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Страница 407: ......