Figure 33 Calibration Setup with Shield
5.
Allow the reading to stabilize and then record to the data collection system for a time
period sufficient to obtain a stable reading.
6.
Repeat steps 1 through 6 for each value of “D” (12’, 14’, 16’, and 18’).
6.4
C
ALCULATING
S
ENSITIVITY
Table 18 explains how to calculate the sensor sensitivity using the Calibration Data sheet. A
sample Calibration Data sheet is provided in Figure 34.
Table 18 Senor Sensitivity Calculation
P+S+A
P Sca Ambient is the reading in amps (NO
shield between the source and the chamber.
S+A
Sca Ambient is the reading in amps taken with the
shield in place.
P
Primary is the difference of the readings:
amps(P) = amps (P+S+A) – amps (S+A)
Exposure rate
Determined for each distance involved from the exposure
rate at one meter from the source.