RSDetection
User’s Manual
Copyright
©
2016 General Electric Company. All Rights Reserved
S131-200-UM Rev A
Page 73
g
Parameter
Support
Comments
ZLD
No
Returns 0.
ZMD
No
Returns 0.
ZHD
No
Returns 0.
Electrometer Gain
RLN
No
Returns 0.
RMN
No
Returns 0.
RHN
No
Returns 0.
RLV
No
Returns 0.
RMV
No
Returns 0.
RHV
No
Returns 0.
General
UID
Yes
Sets or returns the unit address 0-Z except ‘@’.
HEU
Yes
Sets or returns the R to Sv conversion factor..
DTF
Yes
Legacy date format 0 = mm/dd/yy, 1=dd/mm/yy.
WDE
No
Watchdog enable. Returns 1.
COM1
C1B
Yes
Sets or returns the RS=232 port baud rate. Valid values are:
300,600,1200,2400,9600,19200. User must restart the unit after
this command is sent for the changes to go into effect. Make
sure COM1 is available on the unit before sending this
command.
C1P
Yes
Parity for the RS-232 port. Valid values are: (E, O, N). User must
restart the unit after this command is sent for the changes to
go into effect. Make sure COM1 is available on the unit before
sending this command.
C1H
Yes
Handshaking for the RS-232 port. Valid values are:
0 = No handshake, 1=RTS/CTS, > 1 delay in mSec. User must
restart the unit after this command is sent for the changes to
go into effect. Make sure COM1 is available on the unit before
sending this command.
COM2
C2B
Yes
Sets or returns the RS=232 port baud rate. Valid values are:
300,600,1200,2400,9600,19200. User must restart the unit after
this command is sent for the changes to go into effect.
C2P
Yes
Parity for the RS-232 port. Valid values are: (E, O, N). User must