RSDetection
User’s Manual
Copyright
©
2016 General Electric Company. All Rights Reserved
S131-200-UM Rev A
Page 15
g
5.
If the CoreLoad.cab file requires updating, select Upload Unit Firmware from the File menu
and browse to the CD folder (Note: Files can be copied to the computer local drive instead).
Select CoreLoad.cab from the CD path \Firmware\CoreLoad Ver XX.XX.XX\. A progress bar
displays. The file may take several minutes to download.
After the file downloads into the unit, a notification message displays.
6.
If the RSDetection.cab file requires updating, select Upload Unit Firmware from the File menu
and browse to the CD folder. (Note: Files can be copied to the computer local drive instead).
Select RSDetection.cab from the CD path \RSDetection Ver XX.XX.XX\. A progress bar displays.
After the file downloads into the unit, a notification message displays.
7.
After the files download, turn the power off for 10 seconds and then back on. The unit’s LED
blinks normally after the new firmware installation is complete. Note that if CoreLoad.cab was
updated, installation may take several minutes.
THE RSDETECTION UNIT MUST REMAIN POWERED UP DURING THE UPLOAD AND
INSTALLATION PROCESS. IF POWER IS LOST DURING THE PROCESS, THE UNIT MAY
NEED TO BE RETURNED TO THE FACTORY FOR FIRMWARE INSTALLATION.
THE MOST CURRENT UP-TO-DATE RS DETECTION FIRMWARE CAN BE DOWNLOADED
AT
HTTPS://WWW.GEMEASUREMENT.COM/RADIATION-
MEASUREMENT/ENVIRONMENTAL-MONITORING/RSDETECTION-ENVIRONMENTAL-
RADIATION-MONITOR
3.4
M
AINTENANCE AND
S
ERVICE
The RSDetection requires limited maintenance. Only periodic calibration of the unit and
replacement of the optional battery is necessary. In both cases, the unit should be removed from
its customer-provided mounting hardware prior to servicing.
USE APPROPRIATE LIFTING AND HANDLING TECHNIQUES WHEN INSTALLING THE
UNIT.
3.4.1
Calibration
It is recommended the RSDetection is calibrated on a periodic cycle. Reuter-Stokes
recommends calibration cycles determined by the customer’s quality program. A
description of the calibration method used is detailed in the Calibration section of this
manual.