RSDetection
User’s Manual
March 2016
S131-200-UM Rev A
Page 74
g
Parameter
Support
Comments
restart the unit after this command is sent for the changes to
go into effect.
C2H
No
COM3
C3B
Yes
Sets or returns the RS=232 port baud rate. Valid values are:
300,600,1200,2400,9600,19200. User must restart the unit after
this command is sent for the changes to go into effect.
C3P
Yes
Parity for the RS-232 port. Valid values are: (E, O, N). User must
restart the unit after this command is sent for the changes to
go into effect.
C3H
Yes
Handshaking for the RS-232 port. Valid values are:
0 = No handshake, 1=RTS/CTS, > 1 delay in mSec. User must
restart the unit after this command is sent for the changes to
go into effect.
COM4
C4B
Yes
Sets or returns the RS=232 port baud rate. Valid values are:
300,600,1200,2400,9600,19200. User must restart the unit after
this command is sent for the changes to go into effect. Make
sure COM4 is available on the unit before sending this
command.
C4P
Yes
Parity for the RS-232 port. Valid values are: (E, O, N). User must
restart the unit after this command is sent for the changes to
go into effect. Make sure COM4 is available on the unit before
sending this command.
C4H
Yes
Handshaking for the RS-232 port. Valid values are:
0 = No handshake, 1=RTS/CTS, > 1 delay in mSec. User must
restart the unit after this command is sent for the changes to
go into effect. Make sure COM4 is available on the unit before
sending this command.
Alarm Dial
ADN
No
External Serial Display
XDP
Yes
Set returns the serial port used by the display.
0 = No display, 1 = COM1, 2 – COM2, 4 = COM4.
XDU
Yes
External display units 0 = R/h, 1 = Sv/h.