USE APPROPRIATE LIFTING AND HANDLING TECHNIQUES WHEN INSTALLING THE
UNIT.
IT IS THE CUSTOMER’S RESPONSIBILITY TO ENSURE THAT ALL ELECTRICAL
CONNECTIONS (POWER, COMMUNICATIONS EQUIPMENT, ETC.) ARE EITHER OUTDOOR
CERTIFIED OR APPROPRIATELY HOUSED IN NEMA ENCLOSURES.
The recommended installation steps are as follows:
1.
Ensure the RSDetection is powered off.
2.
Mount the RSDetection on the support structure using customer-provided hardware.
3.
Connect desired communication cables (RS-232, Ethernet, USB). Available cables are listed in
Table 21.
CONNECT THE RSDETECTION UNIT TO THE PC BEFORE STARTING THE CONFIG UTILITY
IF USING THE USBB CONNECTION OR USB TO RS 232 ADAPTOR ON THE PC.
4.
Connect external power cable. The RSDetection power wiring is shown in Table 19.
5.
Apply power to the unit.
6.
Verify the following LED sequence occurs:
Off for approximately 5 seconds
Solid red
Solid yellow
Blinking green (or other blinking sequence as indicated in Table 2 LED Operations)
7.
Monitor data until the unit is operating properly. This can be done using the Configuration
Utility windows (the Measurements or Plots tabs are recommended).
Table 2 LED Operations
Color
Pattern
Indication
Red
Solid
Loading Operating System
Yellow
Solid
Loading Application
Green/Yellow Alternating colors every 2
seconds
No Ethernet connection
Green/Red
One blink each per second
An alarm condition exists
Green
Single blink every 2 seconds
Powered from battery, no errors
Green
Double Blink every 2
seconds
Powered from external power, no errors