RSDetection
User’s Manual
March 2016
S131-200-UM Rev A
Page 34
g
Event
Event Description
State
DoseRateVariationAlarm
The dose rate standard
deviation percent
measurement is used
to determine if the
variation in dose rate is
higher than normal. If
the percent exceeds
the alarm value the
event will become
active.
NoAlarm
AlarmActive
HighVoltageHigh
If the high voltage bias
exceeds the alarm limit
specified in the event
configuration this event
will generate an alarm.
NoAlarm
AlarmActive
HighVoltageLow
If the high voltage bias
is lower than the alarm
limit specified in the
event configuration this
event will generate an
alarm.
NoAlarm
AlarmActive
BatteryVoltageHigh
If the battery voltage
exceeds the alarm limit
specified in the event
configuration this event
will generate an alarm.
NoAlarm
AlarmActive
BatteryVoltageLow
If the battery voltage is
lower than the alarm
limit specified in the
event configuration this
event will generate an
alarm.
NoAlarm
AlarmActive
HighChargingCurrent
If the battery charging
current is higher than
the alarm limit
specified in the event
configuration this event
will generate an alarm.
NoAlarm
AlarmActive
HighBatteryDischarge
If the current supplied
by the battery exceeds
the alarm limit
specified in the event
configuration this event
will generate an alarm.
NoAlarm
AlarmActive