WARNING
USE OF CONTROLS OR ADJUSTMENTS, OR PERFORMANCE OF
PROCEDURES OTHER THAN THOSE SPECIFIED HEREIN, MAY RESULT
IN HAZARDOUS RADIATION EXPOSURE.
4.3 Scans Acquired at the Same Tomographic Plane
IEC standard 60601-2-44:2009, clause 203.107 states that you must be warned when scans are
acquired at the same tomographic plane, i.e. same scan location. The need for the warning is to
make users aware of the potential dose that can be given to the patient when acquiring scans at
the same table location.
When acquiring scans in this mode:
•
Utilize the dose information displayed on the
Scan Settings
screen. The dose information
displayed is covered in the next section, CTDI
vol
.
•
An optional DICOM Structured Report (SR) Dose Report is saved in Series 997.
•
Use proper techniques for the application and anatomy you are scanning.
A warning message is posted when [Confirm Settings] is selected for the following scan types:
SmartPrep, Cine or Axial scans with zero table increment (scan interval)
WARNING
THIS SERIES CONTAINS ONE OR MORE GROUPS WITH MULTIPLE SCANS
AT THE SAME TOMOGRAPHIC PLANE, I.E. SAME LOCATION.
DO YOU WANT TO CONTINUE?
After reading the message, if you wish to continue with the scan, click [Continue], otherwise
click [Cancel].
WARNING
PROLONGED EXPOSURE TO X-RAY IN ONE SPOT MAY CAUSE
REDDENING OR RADIATION BURNS. USERS MUST BE AWARE OF THE
TECHNIQUES USED AND EXPOSURE TIME TO ENSURE SAFE OPERATION.
4.4 Geometric Dose Efficiency
A warning message is posted when the Geometric Dose Efficiency in the Z-direction is less than
70%. Geometric Dose Efficiency is a measure of how much of the X-ray beam in the Z-direction
is used by the system.
WARNING
THE FOLLOWING GROUPS IN THE SERIES CONTAINS GEOMETRIC
EFFICIENCY IN THE Z DIRECTION OF LESS THAN 70%: IMAGES 1-12,13-28.
NOTE:
Images 1-12, 13-28 is an example of a location where Geometric Efficiency in Z is
less than 70%.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 3 Safety
61
Содержание Revolution CT
Страница 1: ...Revolution CT User Manual OPERATING DOCUMENTATION 5480385 1EN Revision 1 ...
Страница 2: ...This page left intentionally blank Revolution CT User Manual Direction 5480385 1EN Revision 1 2 ...
Страница 216: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 216 2 Patient Schedule ...
Страница 370: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 370 7 SnapShot Freeze ...
Страница 440: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 440 4 Exam Split ...
Страница 471: ...Illustration 18 VR Settings screen Revolution CT User Manual Direction 5480385 1EN Revision 1 Chapter 16 Reformat 471 ...
Страница 502: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 502 7 Batch Film ...
Страница 674: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 674 1 Operator Messages ...
Страница 680: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 680 1 Units of Measure ...
Страница 688: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 688 1 Lexicon ...
Страница 692: ...Illustration 2 GE XXL Revolution CT User Manual Direction 5480385 1EN Revision 1 692 1 Auto Gating Profiles ...
Страница 694: ...Illustration 4 GE Gen CardioVasc Revolution CT User Manual Direction 5480385 1EN Revision 1 694 1 Auto Gating Profiles ...
Страница 696: ...Illustration 6 GE CACS Revolution CT User Manual Direction 5480385 1EN Revision 1 696 1 Auto Gating Profiles ...
Страница 701: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 701 ...