11. Press Move to Scan.
12. Deliver the breathing and table movement instructions.
13. Press Start Scan.
14. Repeat steps 11–13 to acquire an orthogonal scout scan.
15. Proceed to
.
2.4 Acquire a heart localizer or calcium score scan (optional, but recommended)
Use this procedure to acquire Axial images that can be used to determine the start and end
locations for the heart scan, and to determine the location for the contrast bolus tracking scan.
1. From the
Scan Settings
window, click [Continue].
2. Choose one of the following options:
○
Option 1: Gated localizer for calcium scoring
1. Make sure the scan type is
Cardiac
.
2. For calcium scoring recommended scan settings, see
.
3. Adjust the position of the Graphic Rx region to fully cover the heart.
○
Option 2: Ungated localizer
1. Maker sure the scan type is
Helical
.
2. Recommended settings: 120 kV, 0.5 seconds, 60 mA (100 mA for large patients),
1.375:1 pitch, 55 table speed, 40 mm detector coverage.
3. Position the localizer lines to cover just the heart, approximately from the carina to 2
cm below the apex of the heart.
Illustration 6: Heart Localizer
3. Evaluate the ECG waveform for heart rate stability.
4. Click [Confirm Settings].
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 13 Cardiac
339
Содержание Revolution CT
Страница 1: ...Revolution CT User Manual OPERATING DOCUMENTATION 5480385 1EN Revision 1 ...
Страница 2: ...This page left intentionally blank Revolution CT User Manual Direction 5480385 1EN Revision 1 2 ...
Страница 216: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 216 2 Patient Schedule ...
Страница 370: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 370 7 SnapShot Freeze ...
Страница 440: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 440 4 Exam Split ...
Страница 471: ...Illustration 18 VR Settings screen Revolution CT User Manual Direction 5480385 1EN Revision 1 Chapter 16 Reformat 471 ...
Страница 502: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 502 7 Batch Film ...
Страница 674: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 674 1 Operator Messages ...
Страница 680: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 680 1 Units of Measure ...
Страница 688: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 688 1 Lexicon ...
Страница 692: ...Illustration 2 GE XXL Revolution CT User Manual Direction 5480385 1EN Revision 1 692 1 Auto Gating Profiles ...
Страница 694: ...Illustration 4 GE Gen CardioVasc Revolution CT User Manual Direction 5480385 1EN Revision 1 694 1 Auto Gating Profiles ...
Страница 696: ...Illustration 6 GE CACS Revolution CT User Manual Direction 5480385 1EN Revision 1 696 1 Auto Gating Profiles ...
Страница 701: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 701 ...