Illustration 2: Protocol Category dropdown
3.3 Color Coding for Kids Protocol Selection
Based on the Broselow-Luten Pediatric System, the Color Coding for Kids system was
developed to help you select the correct pediatric CT protocol. The system divides the protocols
into nine color zones based on height and weight, and incrementally increases scan technique
as the patient’s size increases. This arrangement of protocols assists you in reducing the
variations in pediatric protocol selection. If the patient weight is unavailable, a Broselow-Luten
Tape can also be used to obtain the weight based on the length.
Once you select the Pediatric selector, the Pediatric
Protocol List
displays. Next to the name of
the protocol a color icon along with the zone number will be indicated. The notes area of the
protocol selector will display the zone ranges for weight and length, color name and the weight/
color zone number.
NOTE:
Weight-specific protocols are enforced for all anatomical areas except
Head
,
Orbit
,
and
Miscellaneous
. Protocols in the
Head
and
Orbit
categories are usually defined
based on patient age as opposed to patient weight/height.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 5 Pediatrics and Small Patients
127
Содержание Revolution CT
Страница 1: ...Revolution CT User Manual OPERATING DOCUMENTATION 5480385 1EN Revision 1 ...
Страница 2: ...This page left intentionally blank Revolution CT User Manual Direction 5480385 1EN Revision 1 2 ...
Страница 216: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 216 2 Patient Schedule ...
Страница 370: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 370 7 SnapShot Freeze ...
Страница 440: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 440 4 Exam Split ...
Страница 471: ...Illustration 18 VR Settings screen Revolution CT User Manual Direction 5480385 1EN Revision 1 Chapter 16 Reformat 471 ...
Страница 502: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 502 7 Batch Film ...
Страница 674: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 674 1 Operator Messages ...
Страница 680: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 680 1 Units of Measure ...
Страница 688: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 688 1 Lexicon ...
Страница 692: ...Illustration 2 GE XXL Revolution CT User Manual Direction 5480385 1EN Revision 1 692 1 Auto Gating Profiles ...
Страница 694: ...Illustration 4 GE Gen CardioVasc Revolution CT User Manual Direction 5480385 1EN Revision 1 694 1 Auto Gating Profiles ...
Страница 696: ...Illustration 6 GE CACS Revolution CT User Manual Direction 5480385 1EN Revision 1 696 1 Auto Gating Profiles ...
Страница 701: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 701 ...