8.3.5 SmartHelical
SmartHelical is integrated into all GE CT systems. It decreases image noise as measured by
pixel standard deviation.
8.3.6 Advanced Artifact Reduction (AAR)
A filter that significantly reduces streaking artifacts when highly absorbent objects are in the field
of view (example: large shoulders). AAR is automatically enabled as needed.
8.3.7 Adaptive Statistical Iterative Recon (ASiR-V)
The next generation ASiR-V is a reconstruction technology that may allow for reduced mA in the
acquisition of diagnostic images thus, enabling the physician to reduce the patient dose
depending on the clinical task, patient size, anatomical location, and clinical practice.
8.3.8 kV Assist
kV Assist provides the capability to adjust the kV and NI/mA based on the clinical task of the
protocol and patient size to optimize patient dose.
8.3.9 Auto Gating Configuration
Auto Gating uses the scout to evaluate patient attenuation to calculate patient size and a kV/mA
table for selectiing kV, mA, SFOV, and ASiR-V level for the protocol.
8.3.10 Dose Reports
CTDI
vol
, DLP and Dose Efficiency are displayed during the scan prescription and provide patient
dose information. The CTD
Ivol
, DLP, and phantom size used to calculate dose are automatically
saved once you end the exam. The Dose Report is saved as a DICOM secondary capture
image in Series 999. This series can be filmed, archived, and networked after the scan is
completed.
DICOM Structured Dose Report generates a CT Dose Report, which can enable tracking of
dose for the patient by the hospital radiation tracking system or HIS/RIS. The Dose Report is
saved as a DICOM SR Object in Series 997.
8.4 View the Dose Report
CTDI
vol
, DLP, and Dose Efficiency are displayed during the scan prescription to provide patient
dose information.
8.4.1 Prerequisites
Before DICOM Structured Dose Report is turned on, confirm that your PACS or other receiving
station supports structured reports.
Contact your service representative to enable DICOM SR Dose Report.
8.4.2 Dose Report Generation
After closing the exam:
•
The CTDI
vol
, DLP, and Phantom size used to calculated dose are automatically saved.
•
DICOM Structured Dose Report generates a CT Dose Report which can enable tracking of
the patient’s dose information with the hospital radiation tracking system or HIS/RIS.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 11 Scan
305
Содержание Revolution CT
Страница 1: ...Revolution CT User Manual OPERATING DOCUMENTATION 5480385 1EN Revision 1 ...
Страница 2: ...This page left intentionally blank Revolution CT User Manual Direction 5480385 1EN Revision 1 2 ...
Страница 216: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 216 2 Patient Schedule ...
Страница 370: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 370 7 SnapShot Freeze ...
Страница 440: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 440 4 Exam Split ...
Страница 471: ...Illustration 18 VR Settings screen Revolution CT User Manual Direction 5480385 1EN Revision 1 Chapter 16 Reformat 471 ...
Страница 502: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 502 7 Batch Film ...
Страница 674: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 674 1 Operator Messages ...
Страница 680: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 680 1 Units of Measure ...
Страница 688: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 688 1 Lexicon ...
Страница 692: ...Illustration 2 GE XXL Revolution CT User Manual Direction 5480385 1EN Revision 1 692 1 Auto Gating Profiles ...
Страница 694: ...Illustration 4 GE Gen CardioVasc Revolution CT User Manual Direction 5480385 1EN Revision 1 694 1 Auto Gating Profiles ...
Страница 696: ...Illustration 6 GE CACS Revolution CT User Manual Direction 5480385 1EN Revision 1 696 1 Auto Gating Profiles ...
Страница 701: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 701 ...