NOTE:
Flip/Rotate options are only available after service checks that the viewing stations
will display the data correctly.
6.1.9 Copy Forward
Copy Forward Patient Orientation, Patient Position and Anatomical Reference is a feature that
adjusts these three bundled parameters in a scanning protocol based on the values in the
previous series. This feature is located in Protocol Management.
NOTE:
Copy Forward is set at the destination or recipient series.
For example, if Copy Forward was turned “On” in Series 2 of a protocol, and during
scan Patient Orientation, Patient Position and Anatomical Reference were changed
from Head First, Supine, SN to Feet First, Prone, OM for Series 1, these three
settings will be automatically copied for scanning Series 2.
Similar to Copy Patient Orientation / Position, Copy Forward works within the same
patient exam.
The Copy Forward feature can also be used to duplicate (reuse) the following parameters within
series from primary reconstruction to secondary reconstructions:
•
Start Location
•
End Location
•
Time Between Images (Cine scan type only)
•
Interval (Helical scan type only)
•
DFOV
•
R/L Center
•
A/P Center
To activate Copy Forward, type
d
or
D
(duplicate) in the
Parameter
text field in the receiving
reconstruction.
6.2 Primary Recon collection
From the
Scan Settings
window, click the
Primary Recon
collection.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 11 Scan
291
Содержание Revolution CT
Страница 1: ...Revolution CT User Manual OPERATING DOCUMENTATION 5480385 1EN Revision 1 ...
Страница 2: ...This page left intentionally blank Revolution CT User Manual Direction 5480385 1EN Revision 1 2 ...
Страница 216: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 216 2 Patient Schedule ...
Страница 370: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 370 7 SnapShot Freeze ...
Страница 440: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 440 4 Exam Split ...
Страница 471: ...Illustration 18 VR Settings screen Revolution CT User Manual Direction 5480385 1EN Revision 1 Chapter 16 Reformat 471 ...
Страница 502: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 502 7 Batch Film ...
Страница 674: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 674 1 Operator Messages ...
Страница 680: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 680 1 Units of Measure ...
Страница 688: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 688 1 Lexicon ...
Страница 692: ...Illustration 2 GE XXL Revolution CT User Manual Direction 5480385 1EN Revision 1 692 1 Auto Gating Profiles ...
Страница 694: ...Illustration 4 GE Gen CardioVasc Revolution CT User Manual Direction 5480385 1EN Revision 1 694 1 Auto Gating Profiles ...
Страница 696: ...Illustration 6 GE CACS Revolution CT User Manual Direction 5480385 1EN Revision 1 696 1 Auto Gating Profiles ...
Страница 701: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 701 ...