Preventive maintenance
2026419-033E
CIC Pro™
10-9
1.
Configure the leakage tester like the circuit shown below with
GND
switch
OPEN
and polarity switch
NORM
.
2.
Connect probe to an unpainted, non-anodized chassis ground on the unit under
test.
3.
Set the power switch of the device to
ON
.
4.
Read the current leakage indicated on DMM.
NOTE
Center-tapped and non-center-tapped supply circuits produce different
leakage currents and the UL and IEC limits are different.
5.
Set the polarity switch to
RVS
.
6.
Read the current leakage indicated on DMM.
NOTE
If either reading is greater than the appropriate specification below, the
device under test fails. Contact GE Technical Support.
300 µA (0.3 volts on the DMM), and the device under test is powered from
100-120 V/50-60 Hz
300 µA (0.3 volts on the DMM), and the device under test is powered from a
centered-tapped 200-240 V/50-60 Hz, single-phase circuit
500 µA (0.5 volts on the DMM), and the device under test is powered from a
non-center-tapped, 200-240 V/50-60 Hz, single-phase circuit
7.
Set the
GND
switch on the leakage tester to
CLOSED
.
8.
Read the current leakage indicated on DMM.
9.
Set the polarity switch to
RVS
.
10. Read the current leakage indicated on DMM.
100 µA (0.1 volts on the DMM), and the device under test is powered from 100-
240 V/50-60 Hz
Power Cord
Leakage Tester
NORM
RVS
HIGH
LOW
GND
1K
0.015µF
Power Cord
Device
Under
Test
DMM set to measure AC voltage
GND
DMM
Closed
Open
Probe to exposed conductive chassis
10K
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......