Configuration
2026419-033E
CIC Pro™
6-73
NOTE
The CIC Pro center application must be stopped in order to perform a restore.
Because configuration data is automatically updated to match other CIC Pro
centers found running on the care unit, if other CIC Pro centers are left running,
when the restored CIC Pro center application starts back up, some configuration
data would be overwritten by the automatic update.
When you are restoring a CIC Pro center with an existing backup file, the
restored CIC Pro center will be assigned the same unit name and CIC name.
Therefore, you must be careful to edit the unit name while the restored CIC Pro
center is disconnected from the network.
1.
Stop the CIC Pro center application.
a.
From the multi-patient viewer, click
Setup CIC
.
b.
Click the
Service Password
tab.
c.
Type
mms_com
as the password and press
Enter
.
d.
At the Windows command line prompt, type
stop
and press
Enter
.
2.
Disconnect the CIC Pro center from the CARESCAPE Network IX and MC
networks.
3.
Connect a service PC directly to the CIC Pro center and
service interface remotely via the CARESCAPE network IX network on page 4-
7
.
4.
Click
Configuration > Backup/Restore > Unit Defaults
.
5.
Click
Browse
to select a backup file from the service PC/remote computer.
6.
On the
Choose file
window, navigate to the directory where a file from a
previous backup was saved.
7.
Select the backup file, and click
Open
. The backup file location displays in the
Browse
field.
8.
Click
Upload
. The CIC Pro center uploads this backup file. After the restore is
complete, the message,
Unit defaults have been restored
, is displayed.
9.
Click
Logout
to exit the Webmin application.
10. Restart the CIC Pro center application on this CIC Pro center. From the Windows
taskbar, click
Start > Shut Down > Restart
and press
Enter
.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......