6-38
CIC Pro™
2026419-033E
Configuration
Option
Function
Central
Type the central name (up to four characters).
Make sure no other central stations have the same central name in the same care unit.
All other systems on the GE CARESCAPE Network MC network use the Central name to identify this CIC Pro
center system. The central name should be relevant to the location, such as CS2 or SDU3.
Unit
CAUTION
LOSS OF DATA — Once the care unit name is programmed, and the CIC Pro center is placed in use, avoid
changing the care unit name. Changing the care unit name deletes all full disclosure data and the list of
transmitters/transceivers stored on the CIC Pro center. You must re-enter that data after you change the care unit
name.
Select a Unit name from the drop-down list. When entering a Unit name, the following rules apply:
All other systems on the CARESCAPE Network MC network use the care Unit name to identify this CIC Pro
center. The care unit name should be relevant to the location, such as CCU or ICU1.
It is very important to enter the correct care unit name. Be especially careful of the name’s spelling.
If any other CIC Pro centers are intended to have the same care unit name, make sure the care unit names
match exactly.
If the care unit name is not displayed in the Unit list, you must type it in. Then press
Enter
.
Mirror Central
Display
Set up a mirror image view of a primary CIC Pro center.
See Mirror another CIC Pro center on page 6-41.
Waveforms
Set the waveforms you want to display (up to three in addition to the ECG waveform):
ECG1 <From ECG Source>
: This waveform data automatically displays from the default ECG source.
Waveform 2
to
Waveform 4
: Choose to graph subsequent ECG leads.
Printer/Writer
Set the default print devices used to print alarm graphs, manual graphs, and full disclosure or flow volume loops
data.
NOTE
Use a laser printer when the CIC Pro center is interfacing with the Patient Data Server (PDS) or when
printing full disclosure reports or flow volume loops information. A digital writer does not print data under
these conditions.
Full disclosure reports can potentially be very large and take a very long time to print. This can block
alarm graphs sent to that printer. As a result, the printer selection list for full disclosure is different from
the printer selection list for graphing patient data. The printer selection list may not include printers
defined on the CARESCAPE Network IX network.
Laser
: Choose the printer.
DDW
: If a digital writer is connected to the CIC Pro center’s COM2 port, choose
COM2
.
Full Disclosure
: Choose the laser printer.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......