Precautions
WARNING
ELECTRIC SHOCK— To avoid electric shock, the CIC Pro center
and its accessories should not be placed within the patient
environment, which is a volume related to an object (bed, chair,
table, treadmill, etc.) where a patient is intended to be diagnosed,
monitored, or treated.
WARNING
LOSS OF ALARMS — CIC Pro center V5.1 x is in-unit compatible
with CIC Pro center V4.0.7 or later. It is not in-unit compatible with
versions of CIC Pro center V3.1 or earlier and is not in-unit
compatible with any versions of Centralscope. Sharing of the same
care unit name across central stations having incompatible software
versions can result in lost or corrupted telemetry alarm defaults data
and loss of audible alarms.
See Device compatibility on page C-1.
WARNING
LOST OR INTERMITTENT COMMUNICATION — Do not
exceed a maximum of 15 CIC Pro centers in a single logical care
unit.
Attempting simultaneous displays of a patient monitor (bedside or
telemetry) at too many CIC Pro centers may cause lost or
intermittent communication between CIC Pro centers and the
patient monitor. This is evidenced by
NO COMM
or intermittent
communication conditions for the beds.
The maximum CIC Pro centers viewing a patient bedside monitor
can vary depending on bedside monitor capabilities and network
design.
Both hard-wired and telemetry beds are limited in the number of
remote view connections that can be supported.
There are limitations for the device quantity supported by the
CARESCAPE Network. Please contact GE for guidance on
CARESCAPE Network construction.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......