Configuration
2026419-033E
CIC Pro™
6-23
NOTE
If the
NO COMM
is set to
DISABLE
, a visual only indicator will appear in
the multi-patient viewer window at the CIC Pro center. The clinical user will
not receive an audible notification.
DISABLE
is the factory default setting.
If the
NO COMM
is set to
ENABLE
, the clinical user will receive a crisis
level audible notification in addition to a visual
NO COMM
indicator in the
multi-patient viewer window.
When enabled, the system shall provide audible indication for
NO COMM
in no more than 188 seconds from a loss of network communication with a
monitor.
Complete the following procedure to activate the
NO COMM
alarm on the CIC Pro
center:
1.
If you have not already logged onto Webmin,
2.
Click
Configuration > Set Flags
.
NOTE
The
Set Flags Module
screen shows the current set flags settings, which can
be modified.
3.
In the
NO COMM
field on the
Set Flags Module
screen, click
Enable
.
4.
Click
Submit
.
5.
Restart the application.
See Perform safe restart of the CIC Pro center on page 8-
6.
Verify the new settings prior to monitoring patients.
Configure force age setting, if applicable
If the force age setting is
enabled
, a default age range will
not
be displayed in the
Age
field, and the
New Patient
button will be disabled. If the force age setting is
disabled
,
when a patient admit is attempted and no age is selected, a message will prompt the
user to enter the patient age. This setting is usually applicable for pediatric care areas.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......