6-74
CIC Pro™
2026419-033E
Configuration
11. Verify the new settings prior to monitoring patients.
Scenario 2: Add or upgrade a CIC Pro center in a care unit running multiple CIC
Pro centers
When a CIC Pro center is added to or upgraded in a care unit already running other
CIC Pro centers, configuration data common to the care unit is changed in the new or
upgraded CIC Pro center, when it starts up, to match the other CIC Pro centers found
in the care unit network. Configuration data specific only to the new or upgraded CIC
Pro center does not change. Therefore, restore is not useful in this situation. Verify the
new settings prior to monitoring patients.
Scenario 3: Restore multiple CIC Pro centers in a care unit to a previous
configuration
In certain cases, it may be necessary to restore the CIC Pro centers in a care unit to a
previous configuration.
1.
Shut down all but one of the CIC Pro centers in the care unit.
shutdown of the CIC Pro center on page 8-31.
This prevents the restore from being overwritten by associated data from other
CIC Pro centers running on the care unit network.
NOTE
The CIC Pro center application must be stopped in order to perform a
restore. Because configuration data is automatically updated to match other
CIC Pro centers found running in the care unit, if other CIC Pro centers are
left running, when the restored CIC Pro center application starts back up,
some configuration data would be overwritten by the automatic update.
2.
Stop the CIC Pro center application on the one CIC Pro center that was left
running.
a.
From the multi-patient viewer, click
Setup CIC
.
b.
Click the
Service Password
tab.
c.
Type
mms_com
as the password and press
Enter
.
d.
At the Windows command line prompt, type
stop
and press
Enter
.
3.
Disconnect the CIC Pro center from the CARESCAPE Network IX and MC
networks.
4.
Connect a service PC directly to the CIC Pro center and
service interface remotely via the CARESCAPE network IX network on page 4-
7
.
5.
Click
Configuration > Backup/Restore > Unit Defaults
.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......