Example 2:
1.
CICA goes offline and stops collecting full disclosure data on BED1. This causes
CICB to begin collecting full disclosure data on BED1.
NOTE
The master CIC Pro center detects CICA going offline, and switches full
disclosure data collection over to CICB, within a minute of the offline event.
2.
CICA comes back online after exceeding the
Offline Storage
setting time frame.
Because of this, existing BED1 data is deleted from CICA, and the new BED1
data collection begins on CICA. At this point, CICA and CICB are BOTH
collecting full disclosure data on BED1.
3.
The multi-full disclosure rule does not allow more than one CIC Pro center to
collect full disclosure data from the same bed at the same time (only the data that
goes the farthest back in time (CICB) is kept). Because of this, CICB continues
collecting full disclosure data from BED1, and CICA stops.
NOTE
At every wake-up cycle (1 minute interval), the CIC Pro center full
disclosure master scans all CIC Pro centers within the care unit to determine
if more than one CIC Pro center is collecting full disclosure data from the
same bed (this can happen as part of normal operation). If the master detects
this condition, all CIC Pro centers are instructed to stop collecting full
disclosure data from the bed except for the one CIC Pro center with the
oldest data for the bed.
Combo mode (twin-bed rule)
In Combo mode:
1.
Two beds are on the network: one is a hardwired/bedside monitored bed, and the
other is telemetry bed. Both beds share the same name (e.g. BED), but '*' is
appended to the telemetry bed name (e.g. BED and BED*). Both beds represent
the same patient.
NOTE
If the hardwire or telemetry bed was not functioning in Combo mode, these
two beds would be treated as distinct and separate.
2.
Only the oldest bed data is kept. Data for the other bed is deleted.
3.
No matter which data is kept, full disclosure data collection continues, and the
data is stored under the bed name, but without the '*'.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......