Patient Information
• Select
Cancel
to cancel the test for the new patient, and complete the
pending print tasks.
3.
The
Patient Information
screen opens, enter information for the patient.
Enter Patient Information
Use the methods below to enter or update patient data in the
Patient Information
screen:
• Use a barcode reader, see
Update Patient Information with a Barcode Reader on
.
• Do ADT queries, see
Query Orders or ADT for Patient Demographics on page 33
.
• Open a patient record in the
Patients
list, see
Select a Patient from the Patients List
.
• Attach an order, see
.
• Use a keyboard, see
Enter or Edit Patient Information Using the Keyboard on page
Update Patient Information with a Barcode Reader
Using a barcode reader can simplify the entry of patient information and reduce
the chance of introducing errors. When you scan a patient’s barcode, it retrieves
the patient information encoded in the barcode. You can then verify and modify the
information.
You can scan a patient's barcode using an external barcode reader or the internal
barcode reader of the Acquisition Module.
Before you use an external barcode reader, make sure that:
• The BRCD option to use an external barcode reader is activated on the device.
• The barcode reader is connected to the device, and the device is correctly
configured to use the peripheral. For more information, see the
MAC VU360
™
Resting ECG Analysis System Setup and Configuration Manual
.
• You are in the
Live
or
Full Disclosure
ECG screen, or when patient demographics
banner is open.
Before you use the Acquisition Module barcode reader, make sure that:
• The Acquisition Module is connected to the device, and the barcode settings are
configured on the device. For more information, see the
MAC VU360
™
Resting ECG
Analysis System Setup and Configuration Manual
.
• You are in the
Live
or
Full Disclosure
ECG screen, or when patient demographics
banner is open.
An error message displays if you scan a barcode during the acquisition of a Resting
ECG, Rhythm ECG or a Full Disclosure ECG for the patient.
Use the procedure below to scan the patient’s barcode:
32
MAC VU360
™
Resting ECG Analysis System
2088531-370-2