![GE MAC VU360 Скачать руководство пользователя страница 52](http://html2.mh-extra.com/html/ge/mac-vu360/mac-vu360_operators-manual_5075357052.webp)
Recording an ECG or Rhythm
Item
Name
Description
2
Preview of ECG patient
report
Displays the preview of the ECG patient report. If a patient report
contains multiple pages, a part of another page is displayed on
the right-side of the screen. Press the left and right arrows on the
screen to navigate from one page to another.
For more information on the report formats and the standard
layout of the ECG patient report, see
3
Reject
icon
Select the
Reject
icon to return to the live waveform display in the
Acquisition screen.
4
Accept
icon
Select the
Accept
icon to accept the preview of the ECG patient
report and save it in the
Files
list. The accepted ECG patient
report is refreshed and displayed in the same tab for review with
additional options.
5
Live
tab
Select the
Live
tab to return to the live waveform displayed in
the Acquisition screen. The current ECG preview is retained until
you record another ECG, or go back to the preview and accept or
reject it.
6
Full Disclosure
tab
Click the
Full Disclosure
tab from the patient record menu bar
to navigate to the Full Disclosure waveform and application. The
ECG Preview
does not change until you accept or reject the
ECG
Preview
or record the new ECG.
To accept or reject the ECG preview, perform the steps as follows:
1.
Review the patient report and
Hookup Advisor
status.
2.
If the
CRIT
option is enabled on the device, and one or more critical values
are detected during ECG acquisition, a window opens on top of the
ECG
tab
displaying critical value notifications in the order in which they are detected.
3.
Select
Continue
to acknowledge each notification.
If you try to perform other functions such as accessing the
Settings
or
Service
screen prior to accepting or rejecting the ECG, a message displays indicating that
the ECG is not saved, and the preview will be lost if you navigate to the screen.
Select one of the options as follows:
• If you select
Continue
, the preview will be lost.
• If you select
Cancel
, you can proceed to accept or reject the ECG preview.
52
MAC VU360
™
Resting ECG Analysis System
2088531-370-2