Patient Information
If
Then
Multiple ADT records are found on the MUSE
system
The matching records show on the Acquisition
screen, and a message displays in the notification
area that matching ADT data is found.
The maximum of five records display on the
Acquisition screen.
Select the ADT record to search the order using
that selected record's
Patient ID
.
Only one ADT record is found on the MUSE system System searches for the
Patient ID
order found in
that ADT record.
• If multiple orders are found on the system -
The orders show on the
Orders
tab in a filtered
list. Select the order you will to attach to the
patient test.
• If only one order is found on the system - The
order fills in the
Patient Information
screen, if
there is no mismatch with the patient data on
the screen.
• If no orders are found on the system - A
message displays in the notification area that
no matching remote orders are found.
• If remote order query is failed on the system
- A message displays in the notification area
that remote order query has failed.
No ADT records are found on the MUSE system
A message displays in the notification area that no
ADT data/remote orders found.
Remote ADT data query failed on the MUSE
system
A message opens in the notification area that ADT
query has failed.
When local or remote orders are found, and you try to attach the order:
If
Then
The selected order matches the patient
demographics in the
Patient Information
screen
The following warning message opens on the
screen:
Are you sure you want to attach this order to the
current test?
Select
Yes
to attach the order,
No
to cancel the
action.
The selected order does not match the patient
demographics in the
Patient Information
screen
The following warning message opens on the
screen:
Are you sure you want to attach this order to the
current test?
Select
Yes
to attach the order,
No
to cancel the
action.
38
MAC VU360
™
Resting ECG Analysis System
2088531-370-2