Troubleshooting
Error Message
Error Condition
How to troubleshoot:
CA and Client
Certificates have expired
The CA and Client
certificates have expired.
Obtain valid PEM-encoded CA and client
certificates.
CA Certificate is not
installed
The CA certificate is not
installed in the device.
Obtain a PEM-encoded CA certificate.
Client Certificate is not
installed
The client certificate is
not installed in the device.
Obtain a PEM-encoded client certificate.
CA and Client
Certificates are not
installed
The CA and client
certificates are not
installed in the device.
Obtain valid PEM-encoded CA and client
certificates.
FIPS module is not
detected
FIPS module is not
installed.
Contact your GE Healthcare Service
representative to install the FIPS module.
FIPS unrecoverable error
Incompatible firmware
version of FIPS module.
Contact your GE Healthcare Service
representative to replace the FIPS module
with a compatible version.
FIPS initialization failed.
Reconfigure wireless.
If problem persists,
contact GE service.
Installation of the FIPS
option has failed.
Contact your GE Healthcare Service
representative to install the FIPS option.
Table 37: Errors During Network Connection
Error Message
Error Condition
How to troubleshoot:
IP address conflict
The user entered an
invalid IP address.
Enter the correct IP address.
Invalid subnet mask
The user entered an
invalid subnet mask.
Enter the correct subnet mask.
Invalid default gateway
The user entered an
invalid default gateway.
Enter the correct default gateway.
Invalid primary DNS
The user entered an
invalid primary DNS.
Enter the correct primary DNS.
Invalid secondary DNS
The user entered an
invalid secondary DNS.
Enter the correct secondary DNS.
2088531-370-2
MAC VU360
™
Resting ECG Analysis System
129