Report Formats
Item
Description
7
Page number of the report. The page number increments for each page of the rhythm
report.
For the digital rhythm report, the page number is shown as
Page <x> of <y>
.
For the printed rhythm report, the page number is shown as
Page <x>
.
8
The type of report format (rhythm report).
9
The product name.
10
The
Filter
of the ECG waveform (measured in Hz), indicated with ZPD (for High Pass
filter).
NOTE
:
You can change the filter before or during the recording and/or printing of a
rhythm.
For the printed rhythm report, the printing stops and restarts at the newly selected
filter. A gap is shown on the printed rhythm report where the change occurred. Each
time the rhythm printing starts after a change in filter, a calibration pulse is added
for each lead that tells which filter on the printed rhythm.
For the digital rhythm report, the calibration pulse of the last filter selected during
recording is shown on the full report.
11
The
Gain
of the ECG waveform (measured in mm/mV).
NOTE
:
You can change the gain before or during the recording and/or printing of a rhythm.
For the printed rhythm report, the printing stops and restarts at the newly selected
gain. A gap is shown on the printed rhythm report where the change occurred.
Each time the rhythm printing starts after a change in gain, a calibration pulse is
added for each lead that tells the gain on the printed rhythm.
For the digital rhythm report, the calibration pulse of the last gain selected during
recording is shown on the full report.
12
The
Speed
of the ECG waveform (measured in mm/s).
NOTE
:
You can change the speed before or during the printing of a rhythm.
For the printed rhythm report, the printing stops and restarts at the newly selected
speed. A gap is shown on the printed rhythm report where the change occurred.
Each time the rhythm printing starts after a change in speed, a calibration pulse is
added for each lead that tells the speed on the is printed rhythm.
The digital rhythm report is recorded at the configured rhythm speed.
13
The
Leads
.
2088531-370-2
MAC VU360
™
Resting ECG Analysis System
139
Содержание MAC VU360
Страница 7: ...Glossary 167...