Patient Information Fields
Table 51: Patient Information text boxs
Name
Description
Length
Accepted Values
Standard Patient ID:
1 to 16
characters
• A to Z
• a to z
• 0 to 9
• All characters are supported.
Danish Patient ID:
10
characters
Valid values: '0 to 9' and '-'.
The PID must be in the format
ddmmyy-exxg
or
ddmmyyexxg
,
where:
•
dd
= patient day of birth
•
mm
= patient month of birth
•
yy
= patient year of birth
•
e
= patient year of birth is
calculated as follows:
•
e
= 0, 1, 2, 3 or 4, if the
patient year of birth is
between 1900 to 1999.
•
e
= 5, 6, 7, 8 or 9, if the
patient year of birth is
between 2000 to existent
year.
•
xx
= patient place of birth
•
g
= patient gender
• male = odd number
• female = even number
Patient ID
Identification number given to the
patient.
The Patient ID (PID) can be
configured by the administrator
for a specified country
requirement (for example,
Denmark, Sweden, or Norway),
customize text box names, and
add leading zeroes for specified
character lengths.
If the Patient ID does not agree
with the configuration, an error
message opens by the Patient ID
text box.
If the Patient ID agrees with the
configuration, the patient
Date of
Birth
and
Gender
text boxes are
automatically updated.
Swedish Patient ID:
2088531-370-2
MAC VU360
™
Resting ECG Analysis System
151
Содержание MAC VU360
Страница 7: ...Glossary 167...