Recording an ECG or Rhythm
In pre-acquisition mode, the system checks if 10 seconds of ECG data is
available.
• If 10 seconds of ECG data is available, the system records the previous 10
seconds of data for analysis. You cannot stop or cancel the acquisition at this
time.
• If 10 seconds of ECG data is not available, the system continues recording
until 10 seconds of ECG data is obtained. The
Start ECG
icon changes to
Stop
ECG
, and the count of acquisition progress is displayed on the icon until you
have recorded 10 seconds of data. You can cancel the acquisition before
10 seconds of data is recorded. For more information, see
.
In post-acquistion mode, the system begins recording the next 10 seconds of
ECG data for analysis. The
Start ECG
icon changes to
Stop ECG
, and the 10
seconds count of acquisition progress is displayed on the icon. You can cancel
the acquisition before 10 seconds of data is recorded. For more information, see
.
The patient ECG test report begins generating. Based on the print preview mode
configuration and the
Hookup Advisor
status, the recorded ECG patient test
opens in a separate tab.
If ...
Then ...
Print preview mode is configured
as
Always
Print preview mode is configured
as
Yellow
and the Hookup
Advisor Status is
Yellow
or
Red
Print preview mode is configured
as
Red
and the Hookup Advisor
Status is
Red
The ECG patient report preview is displayed in the
ECG
tab for you to accept or reject the report. For more
information on how to accept or reject the report, see
Accept or Reject an ECG Patient Report on page 50
2088531-370-2
MAC VU360
™
Resting ECG Analysis System
49
Содержание MAC VU360
Страница 7: ...Glossary 167...