Patient Information
1.
Start a test for a new patient. For more information, see
.
2.
Scan the patient’s barcode to populate the
Patient Information
screen.
Position the barcode reader 10 cm to 15 cm (4 inches to 6 inches) above the
barcode to be scanned.
If you are using the Acquisition Module barcode reader, press the
Barcode Scan
button.
If you are using an external barcode reader, the barcode is automatically
scanned.
The barcode reader emits a sound to confirm the barcode scan. The
Patient
Information
screen expands and displays the fields populated with patient
information.
3.
If there is a mismatch between the data scanned from the barcode and existing
patient information, a message displays. Perform one of the steps as follows:
• Select
Use Scanned Data
to populate the barcode data into the relevant
fields on the
Patient Information
screen, and confirm the data entered from
the barcode is accurate.
• Select
Use Current Data
to retain the manually entered information on the
Patient Information
screen, and enter or modify patient information as
necessary.
Query Orders or ADT for Patient Demographics
Make sure that:
• You have privileges to see orders and do remote patient query.
• Use orders of ADT data as configured on the device to do remote patient query.
• A barcode reader is connected to the device.
1.
Start a test for a new patient. For more information, see
.
2088531-370-2
MAC VU360
™
Resting ECG Analysis System
33
Содержание MAC VU360
Страница 7: ...Glossary 167...