Arrhythmia Mode Recording
Function Key
Option
Description
F6
More
Toggles through the softkey options.
F1
Pace Enhance
Standardizes the pace spike. Options are
On
and
Off
.
F3
Patient Data
Opens the
Patient Data Entry
window.
F5
Main Menu
Exits the
Arrhythmia
function and returns to
the
Main Menu
.
Printing Options
When printing an arrhythmia report, you have the following options:
Function
Key
Option
Description
F1
Print Summary
Prints a combined report that includes both
the
Table
and
Episode
formats.
F2
Print Table
Prints a breakdown of the recording in tabular
format. The report includes:
• the analysis duration in minutes and
seconds
• the artifact duration in minutes and seconds
• a code for each event type recorded
• the number of each event type recorded
For a description of the possible event codes,
refer to
“Arrhythmia Codes” on page 68
.
F3
Print Episodes
Prints a standard waveform report of the
recorded events. The signal from all recorded
leads is printed, and each event is marked with
the corresponding arrhythmia code.
For a description of the possible event codes,
refer to
“Arrhythmia Codes” on page 68
.
Arrhythmia Codes
The following table identifies the codes used on the arrhythmia reports and the
events they represent.
Code
Arrhythmia Event
A
Artifact
ASYSTO
Asystole, limit value 3s
CPLT
Ventricular couplet (2 PVCs)
ESC
Ventricular escape beat
L
Learn phase
PAU1
Pause of 1 missed beat
68
MAC™ 800
2060026-001C
Содержание MAC 800
Страница 8: ...8 MAC 800 2060026 001C ...
Страница 26: ...Introduction 26 MAC 800 2060026 001C ...
Страница 64: ...Recording a Resting ECG 64 MAC 800 2060026 001C ...
Страница 70: ...Arrhythmia Mode Recording 70 MAC 800 2060026 001C ...
Страница 76: ...RR Analysis 76 MAC 800 2060026 001C ...
Страница 122: ...System Configuration 122 MAC 800 2060026 001C ...
Страница 128: ...Maintenance 128 MAC 800 2060026 001C ...
Страница 138: ...Creating Barcodes 138 MAC 800 2060026 001C ...
Страница 153: ......