System Configuration
5.
Repeat step
through step
for each saved configuration file you want to
export.
6.
When you are done, press
F6
(
Return
).
Import Setup
The
Import Setup
utility allows you to import up to five system setup files from another
device that were exported to an SD card. This feature is useful to sites with multiple
systems that need to have the same or similar setups.
1.
Insert the SD card with the saved setup file.
2.
From the
Main Menu
, press
F4
(
System Configuration
) >
F6
(
More
) >
F6
(
More
) >
F2
(
Import
).
The
Select Setup for Import
window opens. All saved settings on the device
are listed in the left column. All saved settings on the SD card are listed in the
right column.
3.
In the right pane, select the setup file you want to import.
4.
Press
F1
(
Import
).
The selected file is copied to the device and is displayed in the left column.
5.
Repeat step
through step
for each saved configuration file you want to
import.
6.
When you are done, press
F6
(
Return
).
Exporting Audit Trail
The
Export Audit
function copies the system audit trail in XML format to an SD card.
The Audit Trail tracks the creation, transmission, and deletion of records, changes to
the system setup, and tracks the ID of the users who made each change.
Audit trail log files are saved to the
audittrail
directory on the SD card. Their filenames
are in the format
audittrail_x.log
, where x is a number. When a log file is saved
to the SD card, the system determines whether the card already contains an audit trail
log file and names the new file accordingly. For example, if the card does not contain a
log file, the new file is named
audittrail_0.log
; subsequent files are increments
of 1:
audittrail_1.log
,
audittrail_2.log
,
audittrail_3.log
, and
so on.
After the log file is exported to the SD card, it is cleared from the MAC system.
GE Healthcare recommends exporting the audit trail weekly to long term storage to
meet archive requirements. If the audit trail is not exported regularly, it consumes
storage space and reduces the number of ECGs that you can store on the system.
To export an audit trail, the following conditions must be met:
•
High Security Mode
must be enabled.
See
“Basic Setup” on page 87
.
•
Audit Trail
must be enabled.
See
“Basic Setup” on page 87
.
• The user must have the
Edit Setup
and
Delete Records
permissions set.
See
“User Setup” on page 113
120
MAC™ 800
2060026-001C
Содержание MAC 800
Страница 8: ...8 MAC 800 2060026 001C ...
Страница 26: ...Introduction 26 MAC 800 2060026 001C ...
Страница 64: ...Recording a Resting ECG 64 MAC 800 2060026 001C ...
Страница 70: ...Arrhythmia Mode Recording 70 MAC 800 2060026 001C ...
Страница 76: ...RR Analysis 76 MAC 800 2060026 001C ...
Страница 122: ...System Configuration 122 MAC 800 2060026 001C ...
Страница 128: ...Maintenance 128 MAC 800 2060026 001C ...
Страница 138: ...Creating Barcodes 138 MAC 800 2060026 001C ...
Страница 153: ......