Recording a Resting ECG
the setting of the field:
Start rhythm report on a new page
. This field is located on the
Resting ECG Setup
window. Refer to
“Resting ECG Setup” on page 91
for details.
ECG Reanalysis
You can reanalyze ECGs if the following conditions are met:
• The
Audit Trail
option is disabled on the
Basic Setup
window
• Either the
Measurement and 12SL Interpretation
system option (MI12) or the
Measurement 12SL
system option (ME12) is enabled
•
Reanalysis
is selected on the
Resting ECG Setup
window
Reanalysis allows you to modify the
Global Measurements
fiducial points on acquired
waveforms. Refer to the
Marquette 12SL ECG Analysis Program Physician’s Guide
for
additional information.
Reanalyzing an ECG
Use the following procedure to reanalyze a resting ECG.
For additional information, refer to
“Reanalyzing Layout” on page 59
and
.
1.
After acquiring an ECG, press
More
>
More
>
Reanalyze
.
For instructions on acquiring a resting ECG, refer to
on page 52
2.
Review the waveforms to determine the accuracy of the system-selected
fiducial points.
For a better view of individual waveforms, press
Leads
to toggle through the
waveforms. Refer to
“Reanalyzing Layout” on page 59
for more information.
3.
After you have analyzed the waveforms, use the following procedure to adjust
the fiducial points:
a.
Press
F1
(
Next
) to toggle through the fiducial points.
The selected point changes size and is highlighted green.
b.
When the correct point is selected, use the
trimpad
to adjust its position.
c.
To verify correct positioning, refer to the values in the
Measurement
Legend
in the lower left corner of the display.
For more information on the
Measurement Legend
, refer to
d.
Repeat step
through step
for each fiducial point you want to adjust.
4.
When you are done adjusting the fiducial points, do one of the following:
• Discard your adjustments and start over.
a.
Press
F4
(
Restore
).
The original readings are restored.
b.
Return to step
to start over.
• Save your adjustments; press
F5
(
Save
).
58
MAC™ 800
2060026-001C
Содержание MAC 800
Страница 8: ...8 MAC 800 2060026 001C ...
Страница 26: ...Introduction 26 MAC 800 2060026 001C ...
Страница 64: ...Recording a Resting ECG 64 MAC 800 2060026 001C ...
Страница 70: ...Arrhythmia Mode Recording 70 MAC 800 2060026 001C ...
Страница 76: ...RR Analysis 76 MAC 800 2060026 001C ...
Страница 122: ...System Configuration 122 MAC 800 2060026 001C ...
Страница 128: ...Maintenance 128 MAC 800 2060026 001C ...
Страница 138: ...Creating Barcodes 138 MAC 800 2060026 001C ...
Страница 153: ......