Troubleshooting
ACI-TIPI Statement is not Included on Report
If the ACI-TIPI statement is not displayed when expected, do the following:
• Verify the
ACI-TIPI
option is activated.
For information on activating the
ACI-TIPI
option, refer to
“Options Setup” on page
.
• Verify
ACI-TIPI
is enabled on the ECG.
For information, refer to
“Resting ECG Setup” on page 91
• Verify the
ACI-TIPI
required information was entered.
The ACI-TIPI statement prints only if the patient’s gender, date of birth, and chest
pain indication are included in the patient information.
• Verify the patient is 16 or older.
The ACI-TIPI statement does not print for pediatric patients.
• Verify the original ECG was acquired in an electrocardiograph with the
ACI-TIPI
option.
If you attempt to print an ECG that was imported from an external device, the
cart does not generate an ACI-TIPI statement; it prints only if the statement was
saved as part of the ECG.
Paper Jams
If the paper jams while printing, verify the paper was inserted correctly. For details,
refer to the
MAC™ 800 v2.0 Supplies and Accessories Reference Manual
.
SD Card not Present
If you receive an error message stating that the SD card is not present or cannot be
found, do the following:
• Verify an SD card is inserted into the card slot on the device.
For details, refer to
“Rear View” on page 29
.
• Verify the SD card is seated firmly.
The SD card clicks into place when seated firmly.
• Verify the SD card is formatted for a
FAT16
or
FAT32
file system.
To verify an SD card is formatted for the correct file system, do the following:
1.
Insert the card into an SD card reader attached to a PC.
2.
Copy any files you want to save from the SD card to a folder on the PC.
3.
Using the Windows
Format
command, specify either
FAT16
or
FAT32
for the file
system and format the card.
NOTE:
Formatting the SD card erases any existing files on the card.
4.
Copy the files from the folder on the PC to the newly formatted SD card.
Cannot Import or Transmit Records via Modem
If you receive an error while attempting to import or transmit ECG records via modem,
do the following:
• Verify the correct communication option was activated.
2060026-001C
MAC™ 800
131
Содержание MAC 800
Страница 8: ...8 MAC 800 2060026 001C ...
Страница 26: ...Introduction 26 MAC 800 2060026 001C ...
Страница 64: ...Recording a Resting ECG 64 MAC 800 2060026 001C ...
Страница 70: ...Arrhythmia Mode Recording 70 MAC 800 2060026 001C ...
Страница 76: ...RR Analysis 76 MAC 800 2060026 001C ...
Страница 122: ...System Configuration 122 MAC 800 2060026 001C ...
Страница 128: ...Maintenance 128 MAC 800 2060026 001C ...
Страница 138: ...Creating Barcodes 138 MAC 800 2060026 001C ...
Страница 153: ......