Preparing the Patient
Frank X, Y, Z Placement
AHA Label
IEC Label
Description
1
LA black
L yellow
Just below the clavicle of the left arm
2
E orange
E light blue
Mid-sternum on the same horizontal level as 3 and 4
3
V4 blue
C4 brown
Mid-clavicular line in the fifth intercostal space
4
V6 purple
C6 purple
Mid-axillary line on the same horizontal level as 3
5
LL red
F green
Left leg, lower abdominal quadrant
6
RL green
N black
Right leg, lower abdominal quadrant
7
I orange
I light blue
Right mid-axillary line on the same horizontal level as 3
and 4
8
RA white
R red
Just below the clavicle of the right arm
9
H orange
H light blue
Back of neck, avoid the carotid artery and jugular vein
10
M orange
M light blue
Center of spine on the same horizontal level as 3 and 4