Troubleshooting
The system supports two options for communicating via modem:
MODC
(for
communicating with a CardioSoft system) and
MODM
(for communicating with a
MUSE system). For more information, refer to
“Options Setup” on page 115
.
• Verify the modem is connected to an analog telephone line using a standard
RJ11 phone jack.
For more information, refer to
“Rear View” on page 29
.
• Verify the device is plugged into an AC outlet and that the device is powered on.
• Check
Communications Setup
to verify the following:
• The correct modem type is selected.
• The correct dialing method is selected and configured accurately.
For details, refer to
“Communication Setup” on page 100
.
• If transmitting records, check the selected location to verify the following:
•
Modem
is the selected device.
• The
Phone Number
is correct.
• The correct
Protocol
is selected.
For details, refer to
“Communication Setup” on page 100
Cannot Export to Shared Directories
To resolve errors received while attempting to export ECG records to a shared
directory, do the following:
• Verify the
LANC
or
WIFC
communication option was activated.
Refer to
“Options Setup” on page 115
for information on activating options.
• Verify connectivity by doing the following:
• If wired LAN is used, verify that the network cables are connected.
• If wireless LAN is used, verify WIFI module is inserted to USB port of the device,
and the wireless networking setting is configured properly.
Refer to
“Communication Setup” on page 100
for instructions on settings.
• Verify the IP, netmask, gateway, and DNS server addresses are all correct.
Refer to
“Communication Setup” on page 100
for instructions on settings.
• Ping the device from the file server to verify that the two devices can
communicate.
• Verify the logon information is correct.
Check the user name, password, and domain information. Refer to
for information on the log on information.
• Verify share and directory permissions.
Ensure that the account used to log on to the shared directory has read/write/create
permissions to both the share and the directory.
Refer to Microsoft Windows
®
online help for instructions on how to set user
permissions.
Cannot Export to FTP Server
To resolve errors received while attempting to export ECG records to a FTP server, do
the following:
• Verify the LANC or WIFC communication option was activated.
132
MAC™ 800
2060026-001C
Содержание MAC 800
Страница 8: ...8 MAC 800 2060026 001C ...
Страница 26: ...Introduction 26 MAC 800 2060026 001C ...
Страница 64: ...Recording a Resting ECG 64 MAC 800 2060026 001C ...
Страница 70: ...Arrhythmia Mode Recording 70 MAC 800 2060026 001C ...
Страница 76: ...RR Analysis 76 MAC 800 2060026 001C ...
Страница 122: ...System Configuration 122 MAC 800 2060026 001C ...
Страница 128: ...Maintenance 128 MAC 800 2060026 001C ...
Страница 138: ...Creating Barcodes 138 MAC 800 2060026 001C ...
Страница 153: ......