Basic Service Manual

Part Number: 5496012-100

Revision: 3

Содержание LOGIQ V5

Страница 1: ...GE LOGIQ V5 LOGIQ V3 Basic Service Manual Part Number 5496012 100 Revision 3...

Страница 2: ......

Страница 3: ...ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES QUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS T CONSULT ET COMPRIS LE NON RESPECT DE CET AVERTISSEMENT P...

Страница 4: ...IENTE DEVIDO A CHOQUES EL TRICOS MEC NICOS OU OUTROS ESTE MANUAL DE ASSIST NCIA EST DISPON VEL APENAS EM INGL S SE QUALQUER OUTRO SERVI O DE ASSIST NCIA T CNICA QUE N O A GEHC SOLICITAR ESTES MANUAIS...

Страница 5: ...MATERIAALIA TARVITTAVAN K NN KSEN HANKKIMINEN ON ASIAKKAAN VASTUULLA L YRIT KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT JA YMM RT NYT T M N HUOLTO OHJEEN MIK LI T T VAROITUSTA EI NOUDATETA S...

Страница 6: ...U N HO PERSON LU NEBO PACIENT VLIVEM ELEKTRICK HOP PROUDU RESPEKTIVE VLIVEM K RIZIKU MECHANICK HO PO KOZEN NEBO JIN MU RIZIKU DENNE SERVICEMANUAL FINDES KUN P ENGELSK HVIS EN KUNDES TEKNIKER HAR BRUG...

Страница 7: ...B TI SU EISTAS D L ELEKTROS SM GIO MECHANINI AR KIT PAVOJ DENNE SERVICEH NDBOKEN FINNES BARE P ENGELSK HVIS KUNDENS SERVICELEVERAND R TRENGER ET ANNET SPR K ER DET KUNDENS ANSVAR S RGE FOR OVERSETTELS...

Страница 8: ...MENT AR PUTEA DUCE LA R NIREA DEPANATORULUI OPERATORULUI SAU PACIENTULUI N URMA PERICOLELOR DE ELECTROCUTARE MECANICE SAU DE ALT NATUR OVAJ PRIRU NIK ZA SERVISIRANJE DOSTUPAN JE SAMO NA ENGLESKOM JEZI...

Страница 9: ...NTA PROUZRO ENE STRUJNIM UDAROM MEHANI KIM I DRUGIM OPASNOSTIMA T TO SERVISN PR RU KA JE K DISPOZ CII LEN V ANGLI TINE AK Z KAZN KOV POSKYTOVATE SLU IEB VY ADUJE IN JAZYK AKO ANGLI TINU POSKYTNUTIE PR...

Страница 10: ...VUZUN NG L ZCE DI INDAK B R D LDE OLMASINI STERSE KILAVUZU TERC ME ETT RMEK M TER N N SORUMLULU UNDADIR SERV S KILAVUZUNU OKUYUP ANLAMADAN EK PMANLARA M DAHALE ETMEY N Z BU UYARININ G Z ARDI ED LMES E...

Страница 11: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL ix ZH CN KO...

Страница 12: ...l comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qualified personnel i e GE s field engineers personnel of third party service compani...

Страница 13: ...or basic service software does not extend to or cover this software or related documentation If you are a GE employee or a customer who has entered into such a license agreement with GE to use this pr...

Страница 14: ...28 3 Chapter 8 Replacement Procedures pages 8 1 to 8 6 3 Important Precautions pages i to x 3 Chapter 4 Functional Checks pages 4 1 to 4 32 3 Chapter 9 Replacement Parts pages 9 1 to 9 10 3 Table of C...

Страница 15: ...tant Conventions 1 5 Conventions Used in Book 1 5 Standard Hazard Icons 1 6 Product Icons 1 7 Safety Considerations 1 10 Introduction 1 10 Human Safety 1 10 Mechanical Safety 1 10 Electrical Safety 1...

Страница 16: ...2 Console Environmental Requirements 2 2 Electrical Requirements 2 2 EMI Limitations 2 4 Scan Probe Environmental Requirements 2 5 Facility Needs 2 6 Purchaser Responsibilities 2 6 Required Features 2...

Страница 17: ...Order 3 10 Physical Inspection 3 10 EMI Protection 3 10 Completing the Installation 3 11 Power On Boot Up 3 11 Power Off Shutdown 3 12 Transducer Connection 3 12 System Configuration 3 13 System Spec...

Страница 18: ...r On Boot Up 4 2 Power Off Shutdown 4 3 Archiving and Loading Presets 4 5 Adjusting the Display Monitor 4 6 System Features 4 7 B Mode Checks 4 10 M Mode Controls 4 14 Doppler Mode Checks 4 16 Color F...

Страница 19: ...ON 5496012 100 REVISION 3 BASIC SERVICE MANUAL Table of Contents v CHAPTER 5 Components and Functions Theory Overview 5 1 Purpose of Chapter 5 5 1 Block Diagram 5 2 System Diagram 5 2 Common Service P...

Страница 20: ...5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL vi Table of Contents CHAPTER 6 Service Adjustments Overview 6 1 Purpose of this chapter 6 6 1 Monitor Adjustments 6 2 Adjustments Proce...

Страница 21: ...ng Trouble Data 7 2 Overview 7 2 Collect Vital System Information 7 2 Collect a Trouble Image with Logs 7 3 Screen Captures 7 4 Check and Record the Store Key Function 7 4 Setting the Store Key to Scr...

Страница 22: ...N 3 BASIC SERVICE MANUAL viii Table of Contents CHAPTER 8 Replacement Procedures Overview 8 1 Purpose of Chapter 8 8 1 DISASSEMBLY RE ASSEMBLY 8 2 Warning and Caution 8 2 Returning Shipping for repair...

Страница 23: ...MANUAL Table of Contents ix CHAPTER 9 Renewal Parts Overview 9 1 Purpose of Chapter 9 9 1 List of Abbreviations 9 1 Renewal Parts Lists 9 2 Equipment Models Covered in this Chapter 9 2 Operator Consol...

Страница 24: ...How often should care maintenance tasks be performed 10 2 Tools Required 10 4 Standard GE Tool Kit 10 4 Special Tools Supplies and Equipment 10 6 Input Power 10 7 Cleaning 10 7 Physical Inspection 10...

Страница 25: ...s part of the installation or as required during servicing and periodic maintenance 5 Chapter 5 Components and Functions Theory Contains block diagrams and functional explanations of the electronics 6...

Страница 26: ...Manual s should be fully read and understood before operating the LOGIQ V5 LOGIQ V3 and also kept near the unit for quick reference NOTE Probe information displayed on screen does not necessarily ref...

Страница 27: ......

Страница 28: ...5176499 1 N 1018 3Sc RS biopsy kit 5329137 1 N System and Application Software USB 1019 LOGIQ V Series R1 0 0 System and Application Software USB 5501727 1 Y 1020 LOGIQ V Series R1 0 1 System and App...

Страница 29: ...otential hazards are labeled in one of following ways NOTE Notes provide important information about an item or a procedure Information contained in a NOTE can often save you time or effort DANGER DAN...

Страница 30: ...dition to text several different graphical icons symbols may be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific procedures that...

Страница 31: ...s the degree of protection provided by the enclosure per IEC60 529 Bottom of footswitch Type BF Applied Part man in the box symbol is in accordance with IEC 60878 02 03 Beside the probe connector Gene...

Страница 32: ...nvironmental pollution bodily injury or damage to any assets Probe and Rear Panel China Rating Plate Do not place any objects on the monitor Back of LCD Do not force the LCD monitor with your hands Ba...

Страница 33: ...stration is complete this label will be located on the console rating plate NRTL Listing and Certification Mark is used to designate conformance to nationally recognized product safety standards The M...

Страница 34: ...COME UNSTABLE AND TIP OVER WARNING WARNING ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHE...

Страница 35: ...nt chassis must be connected to an electrical ground The system is equipped with a three conductor AC power cable This must be plugged into an approved electrical outlet with safety ground The power o...

Страница 36: ...G WARNING WARNING EXPLOSION WARNING DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A DEFINITE SAFETY HAZARD WARNING WA...

Страница 37: ...waste disposal regulation Equipment being returned must be clean and free of blood and other infectious substances GE policy states that body fluids must be properly removed from any part or equipmen...

Страница 38: ...equipment 3 Isolate the equipment 4 Apply Lockout Tagout Devices 5 Control all stored and residual energy 6 Verify isolation All potentially hazardous stored or residual energy is relieved NOTICE Ener...

Страница 39: ...power of signal cables In addition to electromagnetic energy EMC also includes possible effects from electrical fields magnetic fields electrostatic discharge and disturbances in the electrical power...

Страница 40: ...or Customer Assistance LOCATION PHONE NUMBER USA GE Medical Systems Ultrasound Service Engineering 9900 Innovation Drive Wauwatosa WI 53226 Service On site Service Parts Application Support 1 800 437...

Страница 41: ...Manufacturer FAX Number GE Medical Systems China Co Ltd No 19 Changjiang Road Wuxi National Hi Tech Development Zone Jiangsu P R China 214028 TEL 86 510 85225888 FAX 86 510 85226688 Table 1 9 Factory...

Страница 42: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 1 18 Section 1 6 Customer Assistance This page was intentionally left blank...

Страница 43: ...2 This chapter provides the information required to plan and prepare for the installation of LOGIQ V5 LOGIQ V3 Included are descriptions of the facility and electrical needs to be met by the purchaser...

Страница 44: ...nd equipment This dedicated power shall originate at the last distribution panel before the system Sites with a mains power system with defined Neutral and Live The dedicated line shall consist of one...

Страница 45: ...es 2 2 2 5 Unit Power Plug If the unit arrives without the power plug or with the wrong plug you must contact your GE dealer or the installation engineer must supply what is locally required 2 2 2 6 P...

Страница 46: ...rounding is the most likely reason a unit will have noisy images Check grounding of the power cord and power outlet Replace all screws RF gaskets covers cores After you finish repairing or updating th...

Страница 47: ...nmental Requirements Operation 10 to 40 C Storage 5 to 50 C NOTE Temperature in degrees C Conversion to Degrees F Degrees C 9 5 32 NOTICE SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERA...

Страница 48: ...s Other connections between pieces of electrical equipment products involved and the accompanying electrical installations are highly sophisticated and special engineering competence is required All e...

Страница 49: ...tem without a defined Neutral The dedicated line shall consist of one phase two lines not shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet...

Страница 50: ...least 762 mm 30 inches 2 Film Viewer 3 Counter Top Sink with hot and cold water and Supplies Storage 4 Linen Supply 5 Probes Supplies 6 Examination Table 1930 x 610 mm 76 x 24 inches 7 Footswitch 8 St...

Страница 51: ...Cabinet 3 Film Viewer 4 Counter Top 5 Counter Top and Sink with hot and cold water 6 Overhead Lights Dimmer Dual Level Lighting bright and dim 7 Emergency Oxygen 8 Suction Line 9 Ultrasound system 10...

Страница 52: ...3 5 Suggested floor plan Ultrasound system and EchoPAC PC in same room Figure 2 2 Suggested Room with EchoPAC PC workstation and Ultrasound Scanner 1 EchoPAC PC workstation parts 2 UPS 3 Ethernet netw...

Страница 53: ...es up the on board monitor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier and at a lower cost 2 3 6 4 DICOM O...

Страница 54: ...s cont d Figure 2 2 Worksheet for DICOM Network Information LOGIQ V5 Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP A...

Страница 55: ...to check and test the unit and external peripherals for electrical safety are included in this procedure Also included LOGIQ V5 LOGIQ V3 in this section are guidelines for transporting the unit to a n...

Страница 56: ...stalled and checked out by an experienced service technician in approximately 1 hour LOGIQ V5 LOGIQ V3 consoles with optional equipment may take slightly longer Table 3 2 Average Installation Time Des...

Страница 57: ...personnel should carry out servicing and troubleshooting NOTE For information regarding packing labels refer to LABELS ON PACKAGE 3 After being transported the unit may be very cold or hot If this is...

Страница 58: ...onnect the unit to a properly grounded power outlet Do not use a three to two prong adapter This defeats safety grounding CAUTION Do NOT wear the ESD wrist strap when you work on live circuits and mor...

Страница 59: ...in Chapter 1 CAUTION Do not lift the unit by the Keyboard Equipment damage may result CAUTION The crate with the LOGIQ V5 LOGIQ V3 weighs approximately 50kg Be prepared for a sudden shift of weight as...

Страница 60: ...and Unpacking the Equipment 4 Open the four plastic locks Note Rotate the inside plastic lock counterclockwise to remove it and then remove the outside lock 5 Remove the outside shipping box 6 Remove...

Страница 61: ...r 3 System Setup 3 7 7 Remove the clear plastic wrapped around the LOGIQ V5 LOGIQ V3 from the unit Note To avoid damaging the unit please use a pair of scissors instead of the knife 8 Remove the acces...

Страница 62: ...packing straps around the four wheels and remove the foams beside the wheels Note To aviod injury lock the wheels before cutting the packing straps 11 Unlock the wheels and then hold the control panel...

Страница 63: ...he front of the unit Before moving store all loose parts in the unit Wrap transducers in soft cloth or foam to prevent damage Although LOGIQ V5 LOGIQ V3 is a mobile machine two people should move it o...

Страница 64: ...IQ V5 LOGIQ V3 is found on a label to the right of the Power switch and External I O on the rear of the system 3 4 2 2 Video Formats Check that the video format is set to the locally used video standa...

Страница 65: ...ed too quickly 3 5 1 1 Scanner Power On When power is applied to the scanner power is distributed to the Cooling Unit Control Panel LCD Peripherals and the Back end Processor 3 5 1 2 Turn on the syste...

Страница 66: ...Set key select Shutdown 4 The shutdown process takes a few seconds and is complete when the power status LED is turned blue 5 Disconnect the probes Clean or disinfect all probes as necessary Store the...

Страница 67: ...5 1 1 25 1 1 inches Table 3 6 Weight of LOGIQ V5 LOGIQ V3 With Monitor and Without Other Peripherals Model Weight kg Weight lb LOGIQ V5 LOGIQ V3 Approximately 45 Approximately 99 2 Table 3 7 Electrica...

Страница 68: ...connected using the side connector panel 3 6 5 1 Rear Panel Connector NOTE The AC printer can be connected to the isolated Printer USB port on the rear panel only WARNING WARNING Equipment damage pos...

Страница 69: ...nter ONLY 3 USB Ports USB Ports 4 Ethernet LAN for InSite Connection RJ45 5 Circuit breaker 6 5A 6 AC Inlet 100 240V Table 3 11 Pin Assignments of External VGA Pin No Signal Pin No Signal 1 RED 9 NC 2...

Страница 70: ...UAL 3 16 Section 3 3 Receiving and Unpacking the Equipment Table 3 14 Pin Assignment of S Video Pin No Output Signal Pin No Output Signal 1 GND 3 Y 2 GND 4 C Table 3 15 Pin Assignment of Composite Vid...

Страница 71: ...nect Peripherals A Connect B W printer to the system Sony UP D711MD Printer can be properly connected using the STD and DC cables Sony UP D711MD is connected to the LOGIQ V5 LOGIQ V3 system via the ST...

Страница 72: ...packing the Equipment 3 6 5 3Connect Peripherals cont d B Connect Sony UP D25MD color printer to the system Sony UP D25MD Color Printer can be properly connected using the isolated Printer USB Port An...

Страница 73: ...connected using the isolated Printer USB Port And connect the power cable of Sony UP D897 Sony UP D898MD to the wall outlet D Connect HP Officejet 100 printer to the system HP Officejet 100 Printer c...

Страница 74: ...ANUAL 3 20 Section 3 3 Receiving and Unpacking the Equipment 3 6 5 3Connect Peripherals cont d E Connect Foot Switch to the system Foot Switch can be properly connected using USB Ports Figure 3 11 Con...

Страница 75: ...1 3 6 5 3Connect Peripherals cont d F Connect the USB Memory to the system The USB Memory can be properly connected using the USB ports G Connect the USB HDD to the system The USB Harddisk can be prop...

Страница 76: ...hlight Standard Print in the Service list Select the printer from the Printer pull down Properties menu For the UP D897 printer select Portrait as orientation Type the printer name in the Name field T...

Страница 77: ...y UP D897 as an example 1 Press Utility System Peripherals Select the UP D897 from the pull down menu under Standard Printer Properties Click Properties 2 Select Properties from Printer pull down menu...

Страница 78: ...like Hospital Department Language Units of measure Date Time and Date Format For information on configuring Software Options Refer to the LOGIQ V5 LOGIQ V3 Basic User Manual Chapter 16 Customizing Yo...

Страница 79: ...Information Services Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Name AE Title Dept LOGIQ SN Type Floor Room REV Comments CONT...

Страница 80: ...5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 3 26 Section 3 9 Loading Base Image Software Section 3 9 Loading Base Image Software Refer to Section 8 3 Loading Base Image Software o...

Страница 81: ...0Software Version check out 3 10 1 Functional Check out 1 Power on LOGIQ V5 LOGIQ V3 scanner and wait until system booting to main screen 2 Press Utility key on control panel 3 Choose the About button...

Страница 82: ...fety and user information is available during the operation and service of the complete system 3 11 1 Product Locator Installation NOTE The Product Locator Installation Card shown may not be same as t...

Страница 83: ...V5 LOGIQ V3 console diagnostics by using the built in service software and power supply Section 4 2 Required Equipment To perform these tests you ll need any of the sector linear or convex transducer...

Страница 84: ...he Cooling Unit Control Panel LCD Peripherals and the Back end Processor 4 3 1 2 Turn on the system Press the Power On Off switch at the front of the system once When the Power On Off switch on the Co...

Страница 85: ...e 2 The System Exit window is displayed 3 Using the Set key select Shutdown 4 The shutdown process takes a few seconds and the power off sequence is complete when the power status LED is turned blue 5...

Страница 86: ...f last boot up The BIOS time is changed by 24 hours earlier than the current system by resetting BIOS time replacing BIOS module or changing BIOS time This warning message is to remind the user to che...

Страница 87: ...chiving Presets to an CD R Disk or USB memory device 1 Insert an empty blank CD R disk into the DVD RW 2 Access to the Utility Menu and select System The Backup sheet will be shown on the LCD display...

Страница 88: ...the DVD RW 2 Access to the Utility Menu and select System The Restore sheet will be shown on the LCD display 3 Select the item to restore either from Resource Files 4 Enter restore destination or brow...

Страница 89: ...Down keys 4 TGC 5 AN Keyboard 6 User Defined Keys 7 Patient Preset key 8 Probe key 9 User Preset key 10 Worksheet key 11 End Exam key 12 Archive key 13 Scan Coach 14 Mode Gain 15 Cursor key 16 Clear...

Страница 90: ...2 100 REVISION 3 BASIC SERVICE MANUAL 4 8 Section 4 3 General Procedure 4 3 5 2 LOGIQ V5 LOGIQ V3 SoftMenu Key Tour 1 Rotary key Rotate to adjust the menu 2 Page Up Down key To turn the menu page up a...

Страница 91: ...en 8 Delete Image 9 Save As Menu 10 Number of Images in Exam 11 Page Indication 12 Touchpad Functionality Status 13 Current date and time Caps Lock lit when on network connection indicator PC connecte...

Страница 92: ...re 4 3 6 B Mode Checks 4 3 6 1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 24 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if i...

Страница 93: ...mizes the spectral data Auto adjusts the Velocity Scale PRF live imaging only baseline shift dynamic range and invert if preset Upon deactivation the spectrum is still optimized Mode Cursor No Display...

Страница 94: ...NOTE You can colorize realtime or CINE images or Timeline CINE Colorizes the gray scale image to enhance the eye s discrimination capability Spectrum Colorize colorizes the spectrum as a function of p...

Страница 95: ...lable in 2D imaging and may be used with any transducer or clinical application when image speckle appears to interfere with the desired image detail LOGIQ View Option No LOGIQ View provides the abili...

Страница 96: ...4 3 7 M Mode Controls 4 3 7 1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 24 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if i...

Страница 97: ...Control Possible Bioeffect Description Benefit Sweep Speed No Changes the speed at which the timeline is swept Available in M Mode Doppler Mode and M Color Flow Mode Anatomical M Mode option Yes Anato...

Страница 98: ...oppler Mode Checks 4 3 8 1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 24 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it is...

Страница 99: ...display Angle Correct No Estimates the flow velocity in a direction at an angle to the Doppler vector by computing the angle between the Doppler vector and the flow to be measured NOTE When the Dopple...

Страница 100: ...e calculation automatically which you select in the Modify Auto Calculation when the system is in a state of freeze or live Simultaneous Duplex Triplex Yes When you select Duplex or Triplex everything...

Страница 101: ...E Color Flow Mode only support by LOGIQ V5 4 3 9 1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 24 in Chapter 3 Installation to the System probe connector 2 Turn...

Страница 102: ...spect to reverse flow or vice versa Baseline adjusts the alias point The default baseline is at the midpoint of the color display and at the midpoint of the color bar reference display Angle Steer Yes...

Страница 103: ...based on the number of reflectors that are moving regardless of their velocity PDI does not map velocity therefore it is not subject to aliasing TVI Option Yes Tissue Velocity Imaging TVI calculates a...

Страница 104: ...to this point but not data entered onto worksheets press Clear 4 3 10 2 Circumference Area Ellipse Measurement 1 Press Measure once an active caliper displays 2 To position the active caliper move th...

Страница 105: ...tes in the currently selected operating mode The probe s default settings for the mode and selected exam are used automatically 4 3 11 3 Deactivating the probe 1 Press the Freeze key 2 Gently wipe the...

Страница 106: ...e the Start Frame button to move to the beginning of the CINE Loop Rotate the Start Frame clockwise to move forward through the CINE Loop Rotate the Start Frame anticlockwise to move backward through...

Страница 107: ...drop down menu 3 Enter the label for the media as shown in Figure 4 16 on page 4 25 It is best to use all capital letters with no spaces or punctuation marks Select Format 4 The system displays a pop...

Страница 108: ...sert a formatted media into the drive 2 Enter enter Utility System Backup Restore NOTE If you are not logged in as GE Service or with administrator privileges the Operator Login window is displayed Lo...

Страница 109: ...ator Login window is displayed Log on with administrator privileges 3 In the Restore list select Patient Archive Report Archive User Defined Configuration and Service 4 In the Media field select the B...

Страница 110: ...hard drive by date Therefore the best way is to label media by date NOTE Images will be moved from the hard drive by date Therefore the best way to label media is by date When images are moved to the...

Страница 111: ...Move the EZBackup EZMove wizard starts 8 Verify the information on the first page of EZBack EZMove wizard then select Next If you want to backup all of the exams in the range even if the exam was prev...

Страница 112: ...data and the storage Select Next to continue NOTE The calculation for the number of backup CD is only an estimate Allow for one additional CD when performing an EZBack EZMove NOTE This message Please...

Страница 113: ...een complete press Next NOTE If you need to insert the next media a message appears providing you with the media label Label the media then insert the next media and press OK 11 When the backup is com...

Страница 114: ...Language settings Desired Language is displayed 4 Check assignment of Printer Keys The default function for Store and Print Keys are Store store image Print print Store and Print Keys can also be assi...

Страница 115: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL Chapter 4 Functional Checks 4 33 This page was intentionally left blank...

Страница 116: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 4 34 Section 4 5 Peripheral Checks...

Страница 117: ...onents and Functions Theory Section 5 1 Overview 5 1 1 Purpose of Chapter 5 This chapter explains LOGIQ V5 LOGIQ V3 system concepts component arrangement and subsystem function Table 5 1 Contents in C...

Страница 118: ...IQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 5 2 Chapter 5 Components and Functions Theory Section 5 2 Block Diagram 5 2 1 System Diagram Figure 5 1 LOGIQ V5 LOGIQ V3 System Di...

Страница 119: ...Functions Theory 5 3 Section 5 3 Common Service Platform 5 3 1 Introduction The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems...

Страница 120: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 5 4 Chapter 5 Components and Functions Theory This page was intentionally left blank...

Страница 121: ...ce Adjustments Section 6 1 Overview 6 1 1 Purpose of this chapter 6 This section describes how to test and adjust the scanner These tests are optional You may use them to check the system for errors T...

Страница 122: ...96012 100 REVISION 3 BASIC SERVICE MANUAL 6 2 Section 6 2 Monitor Adjustments Section 6 2 Monitor Adjustments 6 2 1 Adjustments Procedures To adjust the brightness Press Home and End button on the key...

Страница 123: ...e that help maintain image quality and system operation Very basic host system and board level diagnostics are run whenever power is applied Some Service Tools may be run at the application level Howe...

Страница 124: ...the manufacturer for analysis There are different options to acquire this data that would give different results 7 2 2 Collect Vital System Information The following information is necessary in order...

Страница 125: ...Crash Log Power Supply Temperature NOTE Power Supply and Temperature logs are not currently being updated by the LOGIQ V5 LOGIQ V3 This Alt D function is available at all times When Alt D is pressed...

Страница 126: ...e displayed as shown in Figure 7 2 on page 7 4 If Store key is not set to Whole Screen as shown in Figure 7 2 on page 7 4 proceed to step 5 to record the customer s customized settings 5 In the Destin...

Страница 127: ...vigate to and display the image screen to be captured 2 Press Store This will place a snapshot of the screen on the clipboard displayed at the bottom of the scan image display 3 Press Freeze to unfree...

Страница 128: ...7 3 2 Setting the Store Key to Screen Capture on page 7 4 Refer to Figure 7 2 1 Click Utility on the keyboard 2 Select Connectivity from the Utilities Menu 3 Select the Buttons tab on the Connectivity...

Страница 129: ...ections 7 4 1 1 Disruptive Mode Allows you to enable or disable disruptive mode troubleshooting 7 4 1 2 System Shutdown Allows for system shutdown from the diagnostic menu Select to Restart System or...

Страница 130: ...ith network 2 Enter Utility Connectivity TCP IP in IP settings window check Enable DHCP and select the proper network speed in Network Speed NOTE If user wants to setup static IP address uncheck Enabl...

Страница 131: ...ouble Data 7 9 7 5 1 Network Configuration cont d 3 Select Save and a popup window displays Select OK to restart the system and activate the changes 4 After the system restarts the network icon at the...

Страница 132: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 7 10 Section 7 2 Gathering Trouble Data This page was intentionally left blank...

Страница 133: ...8 1 1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems Table 8 1 Contents in Chapter 8 Section Description Page Number 8 1 Overview 8 1 8 2 D...

Страница 134: ...sk Storage Device In case that any patient information is still residing on the LOGIQ V5 LOGIQ V3 GE will contact the customer and request for urgent collection of that patient information GE will kee...

Страница 135: ...wdrivers 8 2 3 2 Preparations Shut Down the System and disconnect the power cord 8 2 3 3 Removal procedure 1 Unscrew 4 screws at the bottom side and remove the footrest refer to Figure 8 9 on page 8 3...

Страница 136: ...connected during system upgrade 1 Insert the USB disk labeled System Application Software in the USB port of the system 2 Properly turn off the scanner by momentarily pressing the Power On Off Switch...

Страница 137: ...ect one of the options for loading the system Select choice 1 to load the complete disk 7 Press any key to continue when below message display as shown below Figure 8 11 Selection for loading the syst...

Страница 138: ...se Image Software Section 8 3 Loading Base Image Software cont d 8 System will be loaded as shown in the screen below 9 System updating finished refer to Figure 8 14 on page 8 6 10 Remove the USB stic...

Страница 139: ...Assembly Ctrl Control FRU 1 Replacement part available in part hub FRU 2 Replacement part available from the manufacturer lead time involved Int Internal I O Input Output LCD Liquid Crystal Display MO...

Страница 140: ...304 2 Power cord China Class 1 1 AC Power Cord 5176773 2 Power cord India Class 1 1 AC Power Cord 5177195 2 Power cord Argentina Class 1 1 AC Power Cord 5176907 2 Power cord UK Class 1 1 AC Power Cord...

Страница 141: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL Chapter 9 Renewal Parts 9 3 Section 9 4 Operator Console Assy Figure 9 1Operator Console Assy...

Страница 142: ...177123 2 Power cord Europe Class 1 1 Power cord China Class 5176304 2 Power cord China Class 1 1 Power cord India Class 5176773 2 Power cord India Class 1 1 Power cord Argentina Class 5177195 2 Power...

Страница 143: ...LOGIQ V3 User Guide Swedish 1 N 7013 5475843 144 LOGIQ V5 LOGIQ V3 User Guide Korean 1 N 7014 5475843 145 LOGIQ V5 LOGIQ V3 User Guide Russian 1 N 7015 5475843 150 LOGIQ V5 LOGIQ V3 User Guide Polish...

Страница 144: ...100 REVISION 3 BASIC SERVICE MANUAL 9 6 Section 9 3 Renewal Parts Lists 7032 5498971 141 LOGIQ V3 User Guide Chinese 1 N 7033 5498972 141 LOGIQ V5 User Guide Chinese 1 N Table 9 4 LOGIQ V Series Manua...

Страница 145: ......

Страница 146: ...12 RS biopsy kit 5176499 1 N 8018 3Sc RS biopsy kit 5329137 1 N System and Application Software USB 8019 LOGIQ V Series R1 0 0 System and Application Software USB 5501727 1 Y 8020 LOGIQ V Series R1 0...

Страница 147: ...LOGIQV3 Item Part Name Part Number Description Quantity FRU 9000 4C RS 5451471 Probe Center Frequency 3 10MHz 1 1 9001 3Sc RS 47237516 Probe Center Frequency 2 75MHz 1 1 9002 L6 12 RS 5454332 Probe Ce...

Страница 148: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 9 10 Section 9 8 Probes This page was intentionally left blank...

Страница 149: ...iarize yourself with the procedures before performing a task Table 10 1 Contents in Chapter 10 Section Description Page Number 10 1 Overview 10 1 10 2 Why do Maintenance 10 2 10 3 Maintenance Task Sch...

Страница 150: ...on must be documented and maintained on the site Your GE service representative can help you with establishing performing and maintaining records for a quality assurance program Please contact us for...

Страница 151: ...es and Connectors Clean Console Inspect Wheels Casters brakes and Swivel Locks Mobile Unit Check Daily Check Control Panel Movement Mobile Unit Check Daily Console Leakage Current Checks also after co...

Страница 152: ...51 Screwdriver Blade Stubby 9 41322 Pickup Tool Claw type 9 41421 Screwdriver Blade Pocket clip 9 6757 6 pc Needle File Set 9 41594 Screwdriver Blade 1 8 in x 4 in 9 9487 Utility Knife 9 41581 Screwdr...

Страница 153: ...n 9 45381 Pliers Arc Joint 9 1 2 inch 9 44067 Socket 1 1 16 in for 1 2 in drive 9 45092 Pliers Linesman 8 1 2 inch 9 42679 Socket 10MM Hex for 1 2 in drive 2273333 9 42882 Punch Pin 3 32 inch 9 44262...

Страница 154: ...e ground cord Anti Static Vacuum Cleaner 46 194427P278 46 194427P279 120V 230V Air Filter air intake Safety Analyzer The Safety Analyzer tool should be calibrated and compliant with AAMI ESI 1993 or I...

Страница 155: ...they may need to be soaked to remove excess gel Table 10 8 Physical Checks Step Item Description 1 Labeling Verify that all system labeling is present and in readable condition refer to the LOGIQ V5...

Страница 156: ...g the combination of lights that are illuminated Any problems found should be reported to the hospital immediately and the receptacle should not be used NOTE No outlet tester can detect the condition...

Страница 157: ...und wire resistance 1 Turn the LOGIQ V5 LOGIQ V3 unit OFF 2 Plug the unit into the meter and the meter into the tested AC wall outlet 3 Plug the black chassis cable into the meter s CHASSIS connector...

Страница 158: ...e unit ON and OFF with the power line polarity Normal and Reversed Record the highest reading of current When using the Microguard or a similar test instrument its power plug may be inserted into the...

Страница 159: ...ighest leakage current measured 10 4 10 Isolated Patient Lead Sink Leakage Isolation Test reference the procedure in the IEC 60601 1 Select the ALL position on the lead selector Depress the rocker swi...

Страница 160: ...mit GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage Current Test 115V 10uA 10uA 220 240V 500uA 10uA Table 10 10 Maximum Allowance Limit for ECG L...

Страница 161: ...dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from differences in line voltage and test lead placement 10 4 11 3 Meter...

Страница 162: ...ility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OFF before switching them using the POLARITY switch and or the NEUTRAL switch Other...

Страница 163: ...lies with the probe or the scanner NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations...

Страница 164: ...y Scan Format Model Number Serial Number Probe 2 Frequency Scan Format Model Number Serial Number Probe 3 Frequency Scan Format Model Number Serial Number Probe 4 Frequency Scan Format Model Number Se...

Страница 165: ...hassis Source Leakage Current Probe Chassis Source Leakage Current Caster Chassis Source Leakage Current CRT Patient Lead Source Leakage Lead to Ground Patient Lead Source Leakage Lead to Lead Patient...

Страница 166: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 10 18 Section 10 5 When There s Too Much Leakage Current This page was intentionally left blank...

Страница 167: ...n scanner and EchoPAC in same room 2 10 floor plan suggestion scanner and EchoPAC in same room 2 10 Functional Checks 4 1 Basic Measurements 4 22 Control Panel 4 7 Monitor Display 4 9 Peripherals 4 32...

Страница 168: ...SERVICE MANUAL II Index Touch Panel Functional Checks 4 7 Trouble Image with Logs 7 3 Troubleshooting Gathering Trouble Data 7 2 Screen Captures 7 4 Trouble Image with Logs 7 3 Vital System Informati...

Страница 169: ......

Страница 170: ...2014 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA...